Zuijderhoudt Floris M J, Kamphuis Johannes S, Kluitenberg Willem E, Dorresteijn-de Bok Jenneke
Department of Clinical Chemistry, Deventer Ziekenhuis, Deventer, The Netherlands.
Clin Chem Lab Med. 2002 Oct;40(10):1036-9. doi: 10.1515/CCLM.2002.181.
We studied precision and accuracy of a HPLC method for determination of porphyrins in feces. A commercial standard solution appeared to contain less coproporphyrin (15%) than stated by the manufacturer. The between-batch coefficients of variation were often below 15% and were higher than the within-batch coefficients. The precision of porphyrin measurements was not influenced by the type of porphyria. Recoveries of added coproporphyrin and protoporphyrin were 90% and 108%; coefficients of variation were 6% and 19%, respectively.