Corresponding dose response of radiographic film with layered gafchromic film.

作者信息

Butson Martin J, Cheung Tsang, Yu Peter K N

机构信息

Department of Physics and Materials Science, City University of Hong Kong, Kowloon Tong, Hong Kong.

出版信息

Phys Med Biol. 2002 Nov 21;47(22):N285-9. doi: 10.1088/0031-9155/47/22/401.

Abstract

This note investigates the dose response of layered HS Gafchromic film compared to Kodak EDR-2 radiographic film. Using five layers of HS type Gafchromic film a dose response greater than EDR-2 film is achieved at the peak wavelength (0.55 OD/Gy versus 0.3 OD/Gy for EDR-2 film). Even over a broader waveband of 30 nm, which is similar to that found in ultra bright LED scanners, the response was found to be 0.38 OD/Gy as opposed to 0.29 OD/Gy. Measurements averaged over the entire visible spectrum produce a relative dose response of 0.165 OD/Gy for five layer HS and 0.29 OD/Gy for EDR-2 film. Due to this high dose response that is achievable, the five layer HS could be used in applications where small doses are delivered to certain areas and a low dependence of energy response is required for measurement.