Wiezorek Tilo, Schwedas Michael, Scheithauer Marcel, Salz Henning, Bellemann Matthias, Wendt Thomas G
Department of Radiooncology, Friedrich-Schiller-University Jena, Germany.
Strahlenther Onkol. 2002 Dec;178(12):732-6. doi: 10.1007/s00066-002-0986-8.
The use of intensity modulated radiation fields needs an extended quality assurance concept. This consists of a linac related part and a case related part. Case related means the verification of an individual treatment plan, optimized on a CT data set of an individual patient and prepared for the treatment of this patient. This part of the quality assurance work is usually time consuming, delivers only partially quantitative results and is uncomfortable without additional help. It will be shown in this paper how the software VERIDOS will improve the optimization of the case related part of the quality assurance work.
The main function of the software is the quantitative comparison of the calculated dose distribution from the treatment planning software with the measured dose distribution of an irradiated phantom. Several additional functions will be explained. Two self-developed phantoms made of RW3 (solid water) and GAFCHROMIC films or Kodak EDR2 films for the measurement of the dose distributions were used. VERIDOS was tested with the treatment planning systems Helay-TMS and Brainscan.
VERIDOS is a suitable tool for the import of calculated dose matrices from the treatment planning systems Helax-TMS and Brainscan and of measured dose matrices exported from the dosimetry software Mephysto (PTW). The import from other treatment planning systems and scanning software applications for film dosimetry is generally possible. In such case the import function has to be adapted to the special header of the import matrix. All other functions of this software tool like normalization (automatically, manually), working with corrections (ground substraction, factors), overlay/comparison of dose distributions, difference matrix, cutting function (profiles) and export functions work reliable.
VERIDOS improves the optimization of the case related part of the quality assurance work for intensity modulated radiation therapy (IMRT). The diverse functions of the software offer the radiation physicist a wide base to verify the IMRT plan independent from the mode of its delivery (compensator technology or MLC technology).
调强放疗射野的应用需要一个扩展的质量保证概念。这包括与直线加速器相关的部分和与病例相关的部分。与病例相关意味着对基于个体患者CT数据集优化并准备用于该患者治疗的个体化治疗计划进行验证。质量保证工作的这一部分通常很耗时,只能提供部分定量结果,并且在没有额外帮助的情况下操作起来很不方便。本文将展示VERIDOS软件如何改进质量保证工作中与病例相关部分的优化。
该软件的主要功能是将治疗计划软件计算出的剂量分布与照射模体的测量剂量分布进行定量比较。还将解释其他几个功能。使用了两个由RW3(固体水)制成的自行研制的模体以及用于剂量分布测量的GAFCHROMIC薄膜或柯达EDR2薄膜。VERIDOS软件在治疗计划系统Helay - TMS和Brainscan上进行了测试。
VERIDOS是一个适用于从治疗计划系统Helax - TMS和Brainscan导入计算剂量矩阵以及从剂量测定软件Mephysto(PTW)导出的测量剂量矩阵的工具。一般也可以从其他治疗计划系统和用于胶片剂量测定的扫描软件应用程序进行导入。在这种情况下,导入功能必须适应导入矩阵的特殊表头。该软件工具的所有其他功能,如归一化(自动、手动)、校正处理(本底扣除、因子)、剂量分布的叠加/比较、差异矩阵、切割功能(剖面图)和导出功能都运行可靠。
VERIDOS改进了调强放射治疗(IMRT)质量保证工作中与病例相关部分的优化。该软件的多种功能为放射物理学家提供了一个广泛的基础,以独立于其交付方式(补偿器技术或多叶准直器技术)来验证IMRT计划。