血压计的缺陷:血压记录中误差的一个重要来源。

Defects in sphygmomanometers: an important source of error in blood pressure recording.

作者信息

Conceiçao S, Ward M K, Kerr D N

出版信息

Br Med J. 1976 Apr 10;1(6014):886-8. doi: 10.1136/bmj.1.6014.886.

Abstract

Almost half the sphygmomanometers in a teaching hospital group had defects in the control valve which interfered with accurate blood-pressure reading. Ward staff should be taught to check sphygmomanometers regularly and replace control valves; time consumption and cost are low. The cuffs in general use in these hospitals, and of the standard size sold in Britain, had rubber bags which did not encircle the arm of more than half the patients on whom they were used. This deficiency causes over-reading of blood pressure in obese people. The size of error is uncertain but it should be avoided by adopting a cuff with a longer bag.

摘要

在一家教学医院集团中,近一半的血压计存在控制阀缺陷,这影响了血压读数的准确性。应教导病房工作人员定期检查血压计并更换控制阀;耗时和成本都很低。这些医院普遍使用的袖带以及在英国销售的标准尺寸袖带,其橡胶袋不能环绕超过一半使用该袖带的患者的手臂。这种缺陷会导致肥胖人群血压读数偏高。误差大小不确定,但应通过采用袋长更长的袖带来避免。

相似文献

2
Evaluation of normal and large sphygmomanometer cuffs using the Omron 705CP.
J Hum Hypertens. 2001 Feb;15(2):131-4. doi: 10.1038/sj.jhh.1001133.
5
Error in blood-pressure measurement due to incorrect cuff size in obese patients.
Lancet. 1982 Jul 3;2(8288):33-6. doi: 10.1016/s0140-6736(82)91163-1.
6
Obesity, hypertension and home sphygmomanometer cuffs.肥胖、高血压与家庭血压计袖带
Eur J Intern Med. 2010 Aug;21(4):338-40. doi: 10.1016/j.ejim.2010.05.002.
8
[Determination of arm circumference for correct measurement of blood pressure. Results of an intervention study].
Hipertens Riesgo Vasc. 2015 Jan-Mar;32(1):6-11. doi: 10.1016/j.hipert.2014.09.001. Epub 2014 Dec 17.

引用本文的文献

6
Influence of arm position on measurement of blood pressure.手臂位置对血压测量的影响。
Br Med J (Clin Res Ed). 1984 May 26;288(6430):1574-5. doi: 10.1136/bmj.288.6430.1574.
7
Blood pressure measurement: current practice and future trends.血压测量:当前实践与未来趋势。
Br Med J (Clin Res Ed). 1985 Mar 9;290(6470):729-34. doi: 10.1136/bmj.290.6470.729.
10
Sphygmomanometers: errors due to blocked vents.血压计:因通气孔堵塞导致的误差。
Br Med J. 1979 Mar 24;1(6166):789-90. doi: 10.1136/bmj.1.6166.789.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验