Mischke R, Keidel A
Clinic for Small Animals, School of Veterinary Medicine Hannover, Hannover, Germany
Vet J. 2003 Jan;165(1):43-52. doi: 10.1016/s1090-0233(02)00169-7.
The platelet function analyser PFA-100 aspirates blood in vitro from a sample reservoir in disposable test cartridges through a microscopic aperture cut into a biologically active membrane at the end of a capillary. In different cartridges the membrane is coated with collagen and adenosine diphosphate (ADP) or collagen and epinephrine (adrenaline) inducing a platelet plug and closure of the aperture. The closure time and total volume of blood flow through the capillary until closure of its aperture were measured. The correlation between platelet count in samples of thrombocytopenic dogs and results of the collagen/ADP cartridge (closure time: r(S)=-0.579; total volume: r(S)=-0.549) was closer than between platelet count and capillary bleeding time. No significant correlation was observed between platelet count and the results obtained with the collagen/epinephrine cartridge. In addition, a higher sensitivity was obtained for the collagen/ADP cartridge. Injection of acetylsalicylic acid into healthy dogs significantly increased closure time and total volume of both types of cartridges (P<0.01). Two dogs with von Willebrand's disease had abnormal values. In contrast, coagulopathies did not significantly influence the results of the platelet function analyser (P>0.05). Despite adequate sensitivity of measurements using the collagen/ADP cartridge to assess quantitative and qualitative platelet disorders in dogs, the influence of haematocrit (P<0.0001) will limit the clinical application of the analyser.
血小板功能分析仪PFA - 100在体外通过一个位于毛细管末端、切入生物活性膜的微小孔径,从一次性测试盒中的样本储存器吸取血液。在不同的测试盒中,膜涂有胶原蛋白和二磷酸腺苷(ADP)或胶原蛋白和肾上腺素,从而诱导血小板形成凝块并封闭孔径。测量了封闭时间以及在毛细管孔径封闭前通过毛细管的血流总体积。血小板减少症犬样本中的血小板计数与胶原蛋白/ADP测试盒的结果之间的相关性(封闭时间:r(S)= - 0.579;总体积:r(S)= - 0.549)比血小板计数与毛细管出血时间之间的相关性更紧密。未观察到血小板计数与胶原蛋白/肾上腺素测试盒所获结果之间存在显著相关性。此外,胶原蛋白/ADP测试盒具有更高的灵敏度。向健康犬注射乙酰水杨酸会显著增加两种测试盒的封闭时间和总体积(P<0.01)。两只患有血管性血友病的犬的值异常。相比之下,凝血障碍对血小板功能分析仪的结果没有显著影响(P>0.05)。尽管使用胶原蛋白/ADP测试盒进行测量对评估犬的血小板数量和质量异常具有足够的灵敏度,但血细胞比容的影响(P<0.0001)将限制该分析仪的临床应用。