Ong P S, Cox H L
Med Phys. 1976 Mar-Apr;3(2):74-9. doi: 10.1118/1.594207.
A line-focusing x-ray monochromator has been designed and developed to be used with an energy dispersive detection system for the quantitative analysis of trace elements in biological specimens. This instrument uses a cylindrically curved Johansson-type crystal to monochromatize and focus the excitation radiation into a line 0.5 mm wide and 40 mm long. The excitation radiation chosen for these experiments was Mo Kalpha and the total intensity of the line-focused beam was estimated to be 2 X 10(7) cps. Because of the narrow width and high power density of the excitation between beam, small tissue specimens 1 X 1 X 12 mm in size along with a single strand of hair were analyzed for trace elements from Al through Sr. For the transition elements, concentrations of a few parts per million were easily detected with this analyzer.