Suppr超能文献

使用镍钛旋转器械的渐进式与恒定锥度轴设计

Progressive versus constant tapered shaft design using NiTi rotary instruments.

作者信息

Bergmans L, Van Cleynenbreugel J, Beullens M, Wevers M, Van Meerbeek B, Lambrechts P

机构信息

Department of Conservative Dentistry, Leuven BIOMAT Research Cluster, Catholic University of Leuven, Belgium.

出版信息

Int Endod J. 2003 Apr;36(4):288-95. doi: 10.1046/j.1365-2591.2003.00650.x.

Abstract

AIM

To evaluate the influence of a progressive versus constant tapered shaft design on canal preparation by NiTi rotary techniques.

METHODOLOGY

A XMCT-scanner and custom-made software were used to nondestructively analyse the mesial canals of 10 extracted mandibular molars in 3D with a spatial resolution of 12.5 microm. Specimens (n = 10 per group) were scanned before and after preparation using ProTaper (progressive tapered) or K3 (constant tapered) files. Numerical values for volume, curvature, dentine removal and centring ratio were obtained in addition to a visual inspection for canal aberrations. Data were analysed by (multiway factorial) anova, Wilcoxon tests and t-tests.

RESULTS

The volume of total dentine removal (mean +/- SD) was 1.21 +/- 0.66 mm(3) (ProTaper) and 1.06 +/- 0.23 mm(3) (K3) (P > 0.05), and the amount of dentine removal at all separate horizontal regions examined was comparable for both groups. The mean linear dentine removal (transportation) was in the range of 8-212 microm (ProTaper) and 4-187 microm (K3). The resultant centring ratio varied from 0.01 to 0.24 (ProTaper) and from 0.01 to 0.17 (K3), whilst different straightening patterns were observed. A centre displacement towards the furcation coronally was most pronounced for the ProTaper group whereas a centre displacement towards the outer side of the curvature more apically was only observed for the K3 group. No severe canal aberrations were found.

CONCLUSIONS

The progressive tapered shaft design of the ProTaper instrument was less influenced by the mid-root curvature than the constant tapered design of the K3 instrument thereby providing a good centred apical preparation. However, ProTaper instruments tended to transport towards the furcation in the coronal region.

摘要

目的

评估渐进式与恒定锥度轴设计对镍钛旋转技术根管预备的影响。

方法

使用一台XMCT扫描仪和定制软件,以12.5微米的空间分辨率对10颗拔除的下颌磨牙的近中根管进行三维无损分析。使用ProTaper(渐进式锥度)或K3(恒定锥度)锉对样本(每组n = 10)进行预备前后扫描。除了对根管畸形进行目视检查外,还获得了体积、曲率、牙本质去除量和定心率的数值。数据通过(多因素)方差分析、Wilcoxon检验和t检验进行分析。

结果

总牙本质去除量(平均值±标准差)为1.21±0.66立方毫米(ProTaper)和1.06±0.23立方毫米(K3)(P>0.05),两组在所有检查的单独水平区域的牙本质去除量相当。平均线性牙本质去除量(偏移)在8 - 212微米(ProTaper)和4 - 187微米(K3)范围内。所得定心率在0.01至0.24(ProTaper)和0.01至0.17(K3)之间变化,同时观察到不同的拉直模式。ProTaper组向冠方分叉处的中心偏移最为明显,而K3组仅在根尖方向更明显地向曲率外侧发生中心偏移。未发现严重的根管畸形。

结论

与K3器械的恒定锥度设计相比,ProTaper器械的渐进式锥度轴设计受牙根中部曲率的影响较小,从而提供了良好的根尖定心预备。然而,ProTaper器械在冠部区域倾向于向分叉处偏移。