Noninvasive blood pressure measurement in mice using pulsed Doppler ultrasound.

作者信息

Reddy Anilkumar K, Taffet George E, Madala Sridhar, Michael Lloyd H, Entman Mark L, Hartley Craig J

机构信息

Section of Cardiovascular Sciences, Department of Medicine, Baylor College of Medicine, Houston, TX 77030, USA.

出版信息

Ultrasound Med Biol. 2003 Mar;29(3):379-85. doi: 10.1016/s0301-5629(02)00746-9.

Abstract

Existing tail-cuff pressure devices for mice use tail flow sensors that measure only systolic and mean pressure. We developed a method to obtain systolic and diastolic pressure in mice using a pulsed Doppler flow velocity sensor and a tail-cuff and validated the method against pressure signals obtained simultaneously from a fluid-filled catheter. The tail-cuff was pressurized to suprasystolic levels to completely occlude the tail artery and then released gradually. The pressure at which the tail flow reappeared was recorded as systolic and the pressure at which the tail flow became continuous was recorded as diastolic. Regression analysis of tail-cuff pressures over catheter pressures obtained from healthy mice (n = 16) showed a high degree of association (r(sys) = 0.95, r(dia) = 0.94, both at p < 0.001). Bland-Altman analysis showed good agreement between the two methods, with a mean difference of -13 ( +/- 12 SD) mmHg and 3 ( +/- 10 SD) mmHg in the systolic (58 to 250 mmHg) and diastolic (48 to 178 mmHg) pressure measurements, respectively. Bland-Altman plots of tail-cuff blood pressures of a second group of mice (n = 20) showed good agreement between repeated measurements obtained on the same day, but had higher variability between measurements made on different days.

摘要

现有的用于小鼠的尾套压力装置使用的是仅测量收缩压和平均压的尾血流传感器。我们开发了一种利用脉冲多普勒流速传感器和尾套来获取小鼠收缩压和舒张压的方法,并针对从充液导管同时获得的压力信号对该方法进行了验证。将尾套加压至收缩压以上水平以完全阻断尾动脉,然后逐渐释放压力。记录尾血流重新出现时的压力为收缩压,尾血流变为连续时的压力为舒张压。对从健康小鼠(n = 16)获得的尾套压力与导管压力进行回归分析,结果显示二者具有高度相关性(r(sys) = 0.95,r(dia) = 0.94,p均< 0.001)。Bland-Altman分析表明两种方法之间具有良好的一致性,在收缩压(58至250 mmHg)和舒张压(48至178 mmHg)测量中,平均差异分别为-13(±12 SD)mmHg和3(±10 SD)mmHg。第二组小鼠(n = 20)的尾套血压Bland-Altman图显示,同一天进行的重复测量之间具有良好的一致性,但不同天进行的测量之间变异性更高。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验