Suppr超能文献

胸部对除颤电流的阻抗预测。

The prediction of the impedance of the thorax to defibrillating current.

作者信息

Geddes L A, Tacker W A, Schoenlein W, Minton M, Grubbs S, Wilcox P

出版信息

Med Instrum. 1976 May-Jun;10(3):159-62.

PMID:1272091
Abstract

In this paper a technique for predicting thoracic impedance to defibrillator pulses is described. The impedance to low-current (1.0 mA) high-frequency (10-500kHz) sinusoidal current is used as an indicator of the impedance of the thorax to high-current, damped sinusoidal waveform pulses. Results from 71 dogs to which defibrillator shocks of 4 to 220 A peak current were applied show that thoracic impedance can be predicted by this method. This information indicates that it is possible to design a defibrillator that can automatically measure chest impedance prior to a defibrillation shock and deliver a predetermined peak current to the subject.

摘要

本文描述了一种预测胸部对除颤器脉冲阻抗的技术。低电流(1.0 mA)高频(10 - 500kHz)正弦电流的阻抗被用作胸部对高电流、阻尼正弦波形脉冲阻抗的指标。对71只施加了峰值电流为4至220 A除颤电击的狗的实验结果表明,该方法可以预测胸部阻抗。这一信息表明,有可能设计一种除颤器,它能够在除颤电击前自动测量胸部阻抗,并向受试者输送预定的峰值电流。