Rose Alan G, Connelly John H, Park Soon J, Frazier O Howard, Miller Leslie W, Ormaza Sofia
Department of Laboratory Medicine and Pathology, University of Minnesota and Fairview-University Medical Center, Minneapolis, Minnesota 55455, USA.
J Heart Lung Transplant. 2003 May;22(5):594-9. doi: 10.1016/s1053-2498(02)01180-4.
This report describes 2 patients with an aortic bioprosthesis. Both patients developed total thrombotic occlusion of the sub-aortic left ventricular outflow tract consequent to insertion of a left ventricular assist device (LVAD). Replacing a mechanical valve with a bioprosthesis in patients receiving a left ventricular assist device offers no additional protection against thrombosis of the aortic prosthesis. Pericardial patching below the aortic prosthesis at the time of LVAD implantation may be performed, but will significantly impede or prohibit the native ventricle from ejecting blood and demonstrating any degree of recovery.
本报告描述了2例使用主动脉生物瓣的患者。两名患者在植入左心室辅助装置(LVAD)后均发生主动脉下左心室流出道完全血栓性闭塞。在接受左心室辅助装置的患者中,用生物瓣替换机械瓣并不能为主动脉瓣假体血栓形成提供额外保护。在植入LVAD时可在主动脉瓣假体下方进行心包修补,但这将显著阻碍或阻止天然心室射血并显示任何程度的恢复。