Kiemeneij Ferdinand, Fraser Douglas, Slagboom Ton, Laarman GertJan, van der Wieken Ron
Amsterdam Department of Interventional Cardiology, Onze Lieve Vrouwe Gasthuis, Amsterdam, The Netherlands.
Catheter Cardiovasc Interv. 2003 Jun;59(2):161-4. doi: 10.1002/ccd.10444.
Radial artery spasm may cause severe discomfort during radial artery sheath removal. A hydrophilic-coated sheath may reduce the force required to remove a radial sheath. This force may be quantified using an automatic pullback device (APD). The objective of this study was to assess if a hydrophilic coating reduces the required force and discomfort associated with removal of a radial sheath following transradial coronary intervention. Ninety patients undergoing percutaneous coronary intervention via the radial artery were randomly assigned to two groups receiving either coated or uncoated introducer sheaths. Radifocus Introducer II (Terumo) 25 cm, 6 Fr radial sheaths and sheaths that were identical apart from the presence of the coating were used in all patients. The APD was used for sheath removal at the end of the procedure. Three patients (7%) in the coated group experienced discomfort during automatic sheath removal, compared to 12 patients (27%) in the uncoated group (P = 0.02). The maximum pullback force (MPF) was significantly lower in the coated compared to the uncoated group (0.24 +/- 0.31 vs. 0.44 +/- 0.33 kg; P = 0.003). Similarly, the mean pullback force was significantly lower in the coated group (0.14 +/- 0.23 vs. 0.32 +/- 0.24 kg; P < 0.001). Only one patient (2%) in each group had an MPF greater than 1.0 kg together with clinical evidence of radial artery spasm. Removal of the coated Terumo Radifocus sheath requires less force than an identical uncoated sheath. The coated sheath was also associated with less discomfort for the patient.
桡动脉痉挛可能会在拔除桡动脉鞘管时引起严重不适。亲水涂层鞘管可能会减少拔除桡动脉鞘管所需的力量。这种力量可以使用自动回撤装置(APD)进行量化。本研究的目的是评估亲水涂层是否能减少经桡动脉冠状动脉介入术后拔除桡动脉鞘管所需的力量及相关不适。90例经桡动脉进行经皮冠状动脉介入治疗的患者被随机分为两组,分别接受涂层或未涂层的导管鞘。所有患者均使用25 cm、6 Fr的Radifocus Introducer II(泰尔茂)桡动脉鞘管以及除有无涂层外其他均相同的鞘管。在手术结束时使用APD拔除鞘管。涂层组有3例患者(7%)在自动拔除鞘管过程中出现不适,而未涂层组有12例患者(27%)出现不适(P = 0.02)。与未涂层组相比,涂层组的最大回撤力(MPF)显著更低(0.24±0.31 vs. 0.44±0.33 kg;P = 0.003)。同样,涂层组的平均回撤力也显著更低(0.14±0.23 vs. 0.32±0.24 kg;P < 0.001)。每组仅有1例患者(2%)的MPF大于1.0 kg且伴有桡动脉痉挛的临床证据。拔除涂层的泰尔茂Radifocus鞘管所需的力量比相同的未涂层鞘管更小。涂层鞘管给患者带来的不适也更少。