用于速率响应模拟的传感器数据存储与检索的实用性和充分性。

Usefulness and adequacy of sensor data storage and retrieval for rate response simulation.

作者信息

Mahaux V, Waleffe A, Kulbertus H

机构信息

Department of Cardiology, University Hospital Sart Tilman, Liège, Belgium.

出版信息

Pacing Clin Electrophysiol. 1992 Nov;15(11 Pt 1):1688-95. doi: 10.1111/j.1540-8159.1992.tb02956.x.

Abstract

The usefulness of sensor data storage for rate response simulation was evaluated using a new dual chamber rate modulated pacemaker sensitive to acceleration forces (Relay 294-03 [Intermedics Inc.]). The pacemaker can store the sensor output during routine exercise and those values can be used to simulate rate profiles for other rate response settings. The predictive value of this feature was evaluated in three studies (mechanical, external pacemaker, and implanted pacemaker). In the first study, the pacemaker was submitted to three runs of eight different mechanical calibrated to-and-fro movements. In the second study, nine external pacemakers were strapped on healthy volunteers who performed three jogging tests. Finally, the predictive value of the simulation was studied in five implanted patients during three successive walking tests. In each study, the pacemaker was submitted three times to the same activity. The responsiveness was successively set to 5, 1, and 10, and the pacemaker outputs were continuously recorded on a Holter monitor. At the end of the first run, rate profile simulations for slopes 1 and 10 were performed; slope 5 rate response was simulated after the second run. A regression analysis was used to establish the correlation between predicted and achieved pacing rates for each study. The coefficients of correlation between predicted and measured pacing rates for the mechanical, external, and clinical studies were 0.999, 0.985, and 0.823, respectively. The corresponding slopes of regression lines were 1.005, 0.971, and 0.935. Calculated rate profile has a high predictive value and could be used to optimize rate responsive settings without serial exercise testings.

摘要

使用一种对加速力敏感的新型双腔频率应答起搏器(Relay 294-03 [Intermedics公司])评估了传感器数据存储在频率应答模拟中的有用性。该起搏器可在日常运动期间存储传感器输出,这些值可用于模拟其他频率应答设置下的频率曲线。在三项研究(机械、体外起搏器和植入式起搏器)中评估了此功能的预测价值。在第一项研究中,使起搏器经受三次共八种不同的机械校准来回运动。在第二项研究中,将九个体外起搏器绑在健康志愿者身上,这些志愿者进行了三次慢跑测试。最后,在五名植入式起搏器患者进行的三次连续步行测试中研究了模拟的预测价值。在每项研究中,使起搏器三次经受相同的活动。应答性依次设置为5、1和10,并使用动态心电图监测仪连续记录起搏器输出。在第一次运行结束时,对斜率1和10进行频率曲线模拟;在第二次运行后模拟斜率5的频率应答。使用回归分析来确定每项研究中预测起搏频率与实际起搏频率之间的相关性。机械、体外和临床研究中预测起搏频率与实测起搏频率之间的相关系数分别为0.999、0.985和0.823。回归线的相应斜率分别为1.005、0.971和0.935。计算得出的频率曲线具有较高的预测价值,可用于优化频率应答设置,而无需进行系列运动测试。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验