Suppr超能文献

当前单极设备中肌电信号过感知的临床相关性。

The clinical relevance of electromyopotential oversensing in current unipolar devices.

作者信息

Gross J N, Platt S, Ritacco R, Andrews C, Furman S

机构信息

Department of Medicine, Montefiore Medical Center, Bronx, New York 10467.

出版信息

Pacing Clin Electrophysiol. 1992 Nov;15(11 Pt 2):2023-7. doi: 10.1111/j.1540-8159.1992.tb03014.x.

Abstract

UNLABELLED

Electromyopotential oversensing of unipolar pacemakers was first appreciated 20 years ago, but its prevalence in present day devices is less well defined. Thirty-four pacemaker patients, only two with symptoms suggestive of oversensing, were evaluated in unipolar settings to assess the frequency of provocation of oversensing in one or, if present, both pacing channels. The sensing threshold of each patient, whenever possible, was recorded as well.

RESULTS

Atrial oversensing occurred in 11/18 patients (61%), all at sensitivities in the 0.4-1.0 mV range. Ventricular oversensing was noted in 13/33 patients (39%), with all but one programmed to settings of 1.25 mV or more sensitive (i.e. < 1.25 mV). Twenty six of 26 patients amenable to testing had ventricular sensing thresholds of at least 4.0 mV or more. Of the 15 patients amenable to atrial sensing threshold testing, 4 had a threshold of 1.0 mV or < 1.0 mV, 6 had thresholds between 1.0-2.0 mV, and 5 sensed at settings > 2.0 mV.

CONCLUSION

Electromyopotential oversensing remains a relevant issue in current day unipolar pacemakers. Most patients do not describe symptoms related to electromyopotential interference, yet such interference is frequently provoked. Oversensing is common at high sensitivities typically utilized for atrial sensing, but quite unusual at settings necessary for adequate ventricular sensing. Programming unipolar devices to unnecessarily high sensitivities should be avoided or serious consequences may result.

摘要

未标注

单极起搏器的肌电电位过感知在20年前首次被认识到,但目前设备中其发生率尚不清楚。对34例起搏器患者在单极设置下进行评估,以评估一个或两个起搏通道中过感知激发的频率,其中只有2例有提示过感知的症状。尽可能记录每位患者的感知阈值。

结果

11/18例患者(61%)发生心房过感知,所有患者的感知灵敏度均在0.4 - 1.0 mV范围内。13/33例患者(39%)出现心室过感知,除1例患者外,其余患者的程控设置均为1.25 mV或更敏感(即<1.25 mV)。26例适合测试的患者中有26例心室感知阈值至少为4.0 mV或更高。在15例适合进行心房感知阈值测试的患者中,4例阈值为1.0 mV或<1.0 mV,6例阈值在1.0 - 2.0 mV之间,5例在>2.0 mV的设置下可感知。

结论

肌电电位过感知在当今的单极起搏器中仍然是一个相关问题。大多数患者没有描述与肌电电位干扰相关的症状,但这种干扰经常被激发。在通常用于心房感知的高灵敏度设置下,过感知很常见,但在足够的心室感知所需的设置下则相当少见。应避免将单极设备程控到不必要的高灵敏度,否则可能会导致严重后果。

相似文献

2
Unipolar sensing in contemporary pacemakers: using myopotential testing to define optimal sensitivity settings.
J Interv Card Electrophysiol. 1998 Mar;2(1):33-40. doi: 10.1023/a:1009708606035.
9
Bipolar ventricular far-field signals in the atrium.心房中的双极心室远场信号。
Pacing Clin Electrophysiol. 1999 Nov;22(11):1604-13. doi: 10.1111/j.1540-8159.1999.tb00379.x.

引用本文的文献

1
[Guidelines for heart pacemaker therapy].[心脏起搏器治疗指南]
Z Kardiol. 2005 Oct;94(10):704-20. doi: 10.1007/s00392-005-0269-3.
2
Unipolar sensing in contemporary pacemakers: using myopotential testing to define optimal sensitivity settings.
J Interv Card Electrophysiol. 1998 Mar;2(1):33-40. doi: 10.1023/a:1009708606035.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验