Grimbert D, Vecellio L, Delépine P, Attucci S, Boissinot E, Poncin A, Gauthier F, Valat C, Saudubray F, Antonioz P, Diot P
INSERM EMI-U 00-10, IFR 120, Groupes de Pneumologie et Enzymologie, Tours, France.
J Aerosol Med. 2003 Summer;16(2):121-9. doi: 10.1089/089426803321919889.
The purpose of this study was to define nebulization conditions providing delivery of aerosols of EPI-hNE4, an inhibitor of human neutrophil elastase (HNE). EPI-hNE4 was nebulized with Pari LC Star and tested at three concentrations (2.5, 5, and 10 mg/mL). The inhaled mass was measured over 15 min. Particle size distribution was measured by cascade impaction. The effect was also tested of mixing EPI-hNE4 with a (99m)Tc human serum albumin (HSA) tracer on the aerodynamic properties of the aerosol. The inhibitory activity of EPI-hNE4 after nebulization was assessed on purified HNE. The inhaled mass was 32.3 +/- 3.5% (mean +/- SD) after 10 min and 44.2 +/- 3.8% (mean +/- SD) after 15 min. Mass median aerodynamic diameter ranged between 1.2 and 1.8 microm. The (99m)Tc HSA EPI-hNE4 aerosol was similar in terms of particle size distribution (y = 1.0338x - 0.003, r = 0.83). (99m)Tc activity was predictive of EPI-hNE4 mass distribution (y = 1.0278x - 1.6991, r = 0.89). The inhibitory capacity of aerosolized samples remained unchanged after up to 10 min of nebulization. EPI-hNE4 can be nebulized efficiently without decrease in its activity. Mixing this inhibitor with (99m)Tc HSA should allow quantification of its deposition in CF patients.
本研究的目的是确定雾化条件,以实现人中性粒细胞弹性蛋白酶(HNE)抑制剂EPI-hNE4气雾剂的递送。使用Pari LC Star对EPI-hNE4进行雾化,并在三种浓度(2.5、5和10 mg/mL)下进行测试。在15分钟内测量吸入量。通过多级冲击法测量粒径分布。还测试了将EPI-hNE4与(99m)Tc人血清白蛋白(HSA)示踪剂混合对气雾剂空气动力学特性的影响。雾化后EPI-hNE4对纯化的HNE的抑制活性进行了评估。10分钟后吸入量为32.3±3.5%(平均值±标准差),15分钟后为44.2±3.8%(平均值±标准差)。质量中值空气动力学直径在1.2至1.8微米之间。(99m)Tc HSA EPI-hNE4气雾剂在粒径分布方面相似(y = 1.0338x - 0.003,r = 0.83)。(99m)Tc活性可预测EPI-hNE4的质量分布(y = 1.0278x - 1.6991,r = 0.89)。雾化长达10分钟后,雾化样品的抑制能力保持不变。EPI-hNE4可以有效地雾化而不降低其活性。将这种抑制剂与(99m)Tc HSA混合应能对其在CF患者中的沉积进行定量。