Benchmarking beam alignment for a clinical helical tomotherapy device.

作者信息

Balog John, Mackie T R, Pearson D, Hui Susanta, Paliwal Bhudatt, Jeraj Robert

机构信息

TomoTherapy Incorporated, Madison, Wisconsin 53717, USA.

出版信息

Med Phys. 2003 Jun;30(6):1118-27. doi: 10.1118/1.1576395.

Abstract

A clinical helical tomotherapy treatment machine has been installed at the University of Wisconsin Comprehensive Cancer Center. Beam alignment has been finalized and accepted by UW staff. Helical tomotherapy will soon be clinically available to other sites. Clinical physicists who expect to work with this machine will need to be familiar with its unique dosimetric characteristics, and those related to the geometrical beam configuration and its verification are described here. A series of alignment tests and the results are presented. Helical tomotherapy utilizes an array of post-patient xenon-filled megavoltage radiation detectors. These detectors have proved capable of performing some alignment verification tests. That is particularly advantageous because those tests can then be automated and easily performed on an ongoing basis.

摘要

一台临床螺旋断层放射治疗机已安装在威斯康星大学综合癌症中心。射束校准工作已完成并得到威斯康星大学工作人员的认可。螺旋断层放射治疗不久将在其他场所投入临床使用。预计使用这台机器的临床物理学家需要熟悉其独特的剂量学特性,本文将描述与几何射束配置及其验证相关的特性。文中给出了一系列校准测试及其结果。螺旋断层放射治疗利用一系列置于患者身后的充氙气兆伏级辐射探测器。这些探测器已证明能够执行一些校准验证测试。这一点特别有利,因为这些测试随后可以自动化,并易于持续进行。