关于定量评估方法应用的调强放射治疗薄膜剂量学质量保证程序的性能分析

Performance analysis of a film dosimetric quality assurance procedure for IMRT with regard to the employment of quantitative evaluation methods.

作者信息

Winkler Peter, Zurl Brigitte, Guss Helmuth, Kindl Peter, Stuecklschweiger Georg

机构信息

Department of Radiation Oncology, University Medical School, Auenbruggerplatz 32, 8036 Graz, Austria.

出版信息

Phys Med Biol. 2005 Feb 21;50(4):643-54. doi: 10.1088/0031-9155/50/4/006.

Abstract

A system for dosimetric verification of intensity-modulated radiotherapy (IMRT) treatment plans using absolute calibrated radiographic films is presented. At our institution this verification procedure is performed for all IMRT treatment plans prior to patient irradiation. Therefore clinical treatment plans are transferred to a phantom and recalculated. Composite treatment plans are irradiated to a single film. Film density to absolute dose conversion is performed automatically based on a single calibration film. A software application encompassing film calibration, 2D registration of measurement and calculated distributions, image fusion, and a number of visual and quantitative evaluation utilities was developed. The main topic of this paper is a performance analysis for this quality assurance procedure, with regard to the specification of tolerance levels for quantitative evaluations. Spatial and dosimetric precision and accuracy were determined for the entire procedure, comprising all possible sources of error. The overall dosimetric and spatial measurement uncertainties obtained thereby were 1.9% and 0.8 mm respectively. Based on these results, we specified 5% dose difference and 3 mm distance-to-agreement as our tolerance levels for patient-specific quality assurance for IMRT treatments.

摘要

介绍了一种使用绝对校准的射线照相胶片对调强放射治疗(IMRT)治疗计划进行剂量验证的系统。在我们机构,这种验证程序在患者照射前对所有IMRT治疗计划进行。因此,临床治疗计划被转移到体模并重新计算。复合治疗计划照射到单张胶片上。基于单张校准胶片自动进行胶片密度到绝对剂量的转换。开发了一个软件应用程序,包括胶片校准、测量和计算分布的二维配准、图像融合以及一些视觉和定量评估工具。本文的主要主题是对该质量保证程序的性能分析,涉及定量评估的公差水平规范。确定了整个程序(包括所有可能的误差源)的空间和剂量精度及准确性。由此获得的总体剂量和空间测量不确定度分别为1.9%和0.8毫米。基于这些结果,我们指定5%的剂量差异和3毫米的距离一致性作为IMRT治疗患者特定质量保证的公差水平。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验