Tack Denis, De Maertelaer Viviane, Gevenois Pierre Alain
Department of Radiology, Centre Hospitalier Universitaire de Charleroi, Boulevard Janson, 92, B-6000-Charleroi, Belgium.
AJR Am J Roentgenol. 2003 Aug;181(2):331-4. doi: 10.2214/ajr.181.2.1810331.
This study was designed to quantify the radiation dose saved by attenuation-based online tube current modulation applied to multidetector CT (MDCT) of the adult trunk as a function of effective milliampere-second (mAs) presets, sex, and body habitus. SUBJECTS AND METHODS. One hundred twenty patients underwent MDCT of the trunk (60 thoracic, 60 abdominal) with an attenuation-based online tube current modulation. Consecutive acquisitions at standard and two lower effective mAs presets were obtained in each patient. Mean percentage effective mAs reductions were compared for each effective mAs preset, taking into account sex and body mass index.
Mean effective mAs reduction was 16.9% and 20.0% for the chest and the abdomen, respectively. Mean percentage effective mAs reductions were found to be significantly different for sex (chest, p = 0.003; abdomen, p = 0.002) but not significantly different for the different effective mAs presets or body mass index.
Attenuation-based online tube current modulation used with MDCT should be considered as a secondary tool of radiation dose reduction because it saves as much as 20% of the radiation dose on the adult trunk, regardless of initial mAs preset. However, initial decreases of mAs presets by the physician should be considered the primary tool for radiation dose reduction.
本研究旨在量化基于衰减的在线管电流调制应用于成人躯干多排螺旋CT(MDCT)时所节省的辐射剂量,该剂量是有效毫安秒(mAs)预设值、性别和体型的函数。
120例患者接受了基于衰减的在线管电流调制的躯干MDCT检查(60例胸部,60例腹部)。在每位患者中,分别以标准有效mAs预设值和两个较低的有效mAs预设值进行连续扫描。比较每个有效mAs预设值下的平均有效mAs降低百分比,并考虑性别和体重指数。
胸部和腹部的平均有效mAs降低分别为16.9%和20.0%。发现平均有效mAs降低百分比在性别上有显著差异(胸部,p = 0.003;腹部,p = 0.002),但在不同的有效mAs预设值或体重指数方面无显著差异。
基于衰减的在线管电流调制与MDCT联合使用应被视为减少辐射剂量的辅助手段,因为无论初始mAs预设值如何,它都能在成人躯干上节省高达20%的辐射剂量。然而,医生最初降低mAs预设值应被视为减少辐射剂量的主要手段。