Aschendorff Antje, Klenzner Thomas, Richter Bernhard, Kubalek Ralf, Nagursky Heiner, Laszig Roland
Department of Otorhinolaryngology - Head and Neck Surgery, University Clinic Freiburg, Germany.
J Laryngol Otol. 2003 Jul;117(7):527-31. doi: 10.1258/002221503322112932.
The aim of the study presented was to assess the insertion mode and possible intracochlear trauma after implantation of the HiFocus electrode with positioner in human temporal bones. The study was performed in five freshly frozen temporal bones. The position of electrodes was evaluated using conventional X-ray analysis, rotational tomography and histomorphological analysis. Insertion of the HiFocus electrode with positioner resulted in considerable trauma to fine cochlear structures including fracture of the osseous spiral lamina, dislocation of the electrode array from the scala tympani into the scala vestibuli and fracture of the modiolus close to the cochleostomy. The implication of the results regarding clinical outcome will be discussed.
本研究的目的是评估在人类颞骨中植入带定位器的HiFocus电极后的插入方式及可能的耳蜗内创伤。该研究在五块新鲜冷冻的颞骨上进行。使用传统X射线分析、旋转断层扫描和组织形态学分析来评估电极的位置。带定位器的HiFocus电极插入导致对耳蜗精细结构造成相当大的创伤,包括骨螺旋板骨折、电极阵列从鼓阶脱位至前庭阶以及靠近蜗窗的蜗轴骨折。将讨论这些结果对临床结局的影响。