Wadi-Ramahi Shada J, Bernard Damian, Chu James C H
Department of Medical Physics, Rush-Presbyterian-St. Luke's Medical Center, Chicago, Illinois 60612, USA.
Med Phys. 2003 Jul;30(7):1556-65. doi: 10.1118/1.1578484.
Recent work proposed the use of magnetic field as a solution to reduce the undesirable effect of air cavities on dose after the air/tissue interface. In contrast to the published work that looks into the problem with slab geometries, in this work we use actual anatomy based on CT images and the magnetic flux from a Helmholtz coil-pair configuration to investigate the problem and to evaluate the efficacy of the proposed solution. The EGS4 phantom was created using CT scans of the head at the level of the ethmoid sinus. The sinus measures 1.95 x 2.18 x 2.00 cm3. The grid size used is 0.15 x 0.15 x 0.4 cm3. Three different radiation beams, 1 x 1, 2 x 2, and 4 x 4 cm2, all 6 MV irradiate the phantom in two different configurations: single beam and parallel opposed. The magnetic field has three different strengths: 0.0, 0.5, and 1.0 T. These represent the maximum strength achieved in the middle of the configuration, between the two coils. The depth of the second buildup region in the absence of the magnetic field was used as the normalization point for the purpose of analysis. Dose was then scored at 0.23 cm after the air/tissue interface. A second phantom, very similar to the CT-based phantom, was created, but with the sinus cavity filled with unit-density tissue; everything else remained the same. This phantom provides a base to investigate the effect of the air cavity on dose. The phantom was termed the phantom without air, or PWA for short. We use the terms "dose reduction ratio" (DRR), defined as one minus the ratio of the dose in PWA to the dose with the presence of the cavity multiplied by 100% and the "dose improvement ratio" (DIR), defined as the ratio of dose with B to that without B, to evaluate the reduction in dose due to the cavity and the improvement in dose with magnetic field, respectively. For single beam geometry, the reduced dose ranged from 41% (1 x 1 cm2 beam) to less than 2% (4 x 4 cm2 beam). For the same single beam geometry, DIR ranged from 1.13 to 1.00 (DIR = 1 indicates no change) with 0.5 T, whereas it ranged from 1.44 to 1.05 for 1.0 T magnets. When an opposing beam was used, the reduced dose was not as severe, such that DRR ranged from 24% to less than 2%. Whereas the dose improvement ranged from 1.08 to 1.00 for 0.5 T, and from 1.23 to 1.01 for 1.0 T.
最近的研究提出使用磁场来解决空气腔对空气/组织界面后剂量的不良影响。与已发表的关于平板几何形状问题的研究不同,在本研究中,我们使用基于CT图像的实际解剖结构和亥姆霍兹线圈对配置产生的磁通量来研究该问题,并评估所提出解决方案的有效性。EGS4体模是使用筛窦水平的头部CT扫描创建的。该窦的尺寸为1.95×2.18×2.00 cm³。使用的网格尺寸为0.15×0.15×0.4 cm³。三种不同尺寸的辐射束,1×1、2×2和4×4 cm²,均为6 MV,以两种不同配置照射体模:单束和对穿束。磁场有三种不同强度:0.0、0.5和1.0 T。这些代表配置中间、两个线圈之间达到的最大强度。在没有磁场的情况下,第二个建成区的深度用作分析的归一化点。然后在空气/组织界面后0.23 cm处记录剂量。创建了第二个与基于CT的体模非常相似的体模,但窦腔内填充了单位密度组织;其他一切保持不变。这个体模为研究空气腔对剂量的影响提供了一个基础。该体模被称为无空气体模,简称为PWA。我们使用“剂量降低率”(DRR),定义为1减去PWA中的剂量与有腔时的剂量之比再乘以100%,以及“剂量改善率”(DIR),定义为有磁场时的剂量与无磁场时的剂量之比,分别评估由于腔导致的剂量降低和磁场对剂量的改善。对于单束几何形状,剂量降低范围从41%(1×1 cm²束)到小于2%(4×4 cm²束)。对于相同的单束几何形状,0.5 T时DIR范围为1.13至1.00(DIR = 1表示无变化),而1.0 T磁体时DIR范围为1.44至1.05。当使用对穿束时,剂量降低不那么严重,使得DRR范围从24%到小于2%。而0.5 T时剂量改善范围为1.08至1.00,1.0 T时为1.23至1.01。