Huang Zhong-hua, Li Yan, Wang Jun-de, Chen Zuo-ru
Laboratory of Advanced Spectroscopy, Nanjing University of Science and Technology, Nanjing 210014, China.
Guang Pu Xue Yu Guang Pu Fen Xi. 2002 Jun;22(3):399-400.
In order to obtain the infrared spectral radiance distribution of the emission spectrum or the concentration of the gases, the calibration of FTIR emission system is necessary for obtaining an instrument response function of the system. This paper gives the effect of the temperature of the standard blackbody for calibration of EQUINOX55 remote sensing FTIR emission system on the instrument response function. The result illustrates that the values of the instrument response functions increase as the blackbody temperature increased, and tends to a limit.
为了获得发射光谱的红外光谱辐射度分布或气体浓度,对傅里叶变换红外发射系统进行校准对于获得系统的仪器响应函数是必要的。本文给出了用于校准EQUINOX55遥感傅里叶变换红外发射系统的标准黑体温度对仪器响应函数的影响。结果表明,仪器响应函数的值随着黑体温度的升高而增加,并趋于一个极限值。