Lee T, Lee S K, Cheng C H, Liao S Q
Department of Radiology, Veterans General Hospital-Taichung, Taiwan, R.O.C.
Zhonghua Yi Xue Za Zhi (Taipei). 1992 Sep;50(3):198-202.
To evaluate the clinical desired accuracy of the left ventricular ejection fractions (EFs) calculated by magnetic resonance imaging (MRI). Within one to three days post coronary angiography and left ventriculogram, twenty-five adults were studied by MR imaging. They had nuclear medicine studies for left ventricular ejection fractions as well. EKG-gated spin-echo 30-msec echo-delay images were obtained in end systole and end diastole in a plane parallel to the ventricular septum. Analysis of ventricular volumes and ejection fractions were performed using the area-length method. The EFs calculation by gated MRI were compared with that obtained by angiocardiography and nuclear medicine studies. The linear regression line obtained for ejection fraction was y = 0.858x +6.813, r = 0.753, p = 0.009; y' = 1.567x'-24.692, r' = 0.783, p' = 0.002. The above figure indicate a reasonable correlation among these three methods.
评估通过磁共振成像(MRI)计算的左心室射血分数(EFs)的临床期望准确性。在冠状动脉造影和左心室造影后的一至三天内,对25名成年人进行了MR成像研究。他们也进行了左心室射血分数的核医学研究。在与室间隔平行的平面上,于收缩末期和舒张末期获得心电图门控自旋回波30毫秒回波延迟图像。使用面积长度法进行心室容积和射血分数分析。将门控MRI计算的EFs与通过心血管造影和核医学研究获得的结果进行比较。射血分数得到的线性回归线为y = 0.858x +6.813,r = 0.753,p = 0.009;y' = 1.567x'-24.692,r' = 0.783,p' = 0.002。上述数据表明这三种方法之间存在合理的相关性。