Suppr超能文献

通过磁共振成像测定射血分数:与左心室血管造影术的比较。

Ejection fraction determination by MR imaging: comparison with left ventricular angiography.

作者信息

Stratemeier E J, Thompson R, Brady T J, Miller S W, Saini S, Wismer G L, Okada R D, Dinsmore R E

出版信息

Radiology. 1986 Mar;158(3):775-7. doi: 10.1148/radiology.158.3.3945753.

Abstract

To determine the accuracy of left ventricular ejection fractions (EFs) calculated from magnetic resonance (MR) images, 22 patients who underwent coronary angiography and left ventriculography were studied within 1-3 days by MR imaging. ECG-gated spin-echo 30-msec echo-delay images were obtained in end systole and end diastole in a plane through the long axis of the left ventricle perpendicular to the septum at a level through the aortic valve and apex. The area-length method was then used to calculate the EF from left ventriculograms and MR images. Angiographic EF correlated linearly with MR determinations. The mean differences in EF calculation between two observers were 5.0 +/- 6.9 for ventriculography and 2.7 +/- 7.8 for MR imaging. MR imaging, using image planes oriented to the left ventricular axes, can allow accurate EF calculation. This has the potential to be clinically useful in noninvasive evaluation of cardiac function.

摘要

为了确定从磁共振(MR)图像计算出的左心室射血分数(EF)的准确性,对22例接受了冠状动脉造影和左心室造影的患者在1至3天内进行了MR成像研究。在舒张末期和收缩末期,于通过主动脉瓣和心尖的水平,在垂直于室间隔的左心室长轴平面上获得心电图门控自旋回波30毫秒回波延迟图像。然后使用面积-长度法从左心室造影片和MR图像计算EF。血管造影EF与MR测定值呈线性相关。两位观察者之间EF计算的平均差异,心室造影为5.0±6.9,MR成像为2.7±7.8。使用与左心室轴定向的图像平面的MR成像可以实现准确的EF计算。这在心脏功能的无创评估中可能具有临床应用价值。