[Noninvasive evaluation of cardiac index and ejection fraction in ischemic heart disease. Comparison of thoracic bioimpedance, thermodilution and cineventriculography].

作者信息

Talarico G, Saino A, Lotto A

机构信息

Ospedale Maggiore IRCCS, Divisione di Cardiologia, Università degli Studi di Milano.

出版信息

Minerva Cardioangiol. 1992 Jun;40(6):203-9.

PMID:1407614
Abstract

Aim of the study was to verify the reliability of thoracic bioimpedance cardiography (TEB) in detection, non-invasively, cardiac index (IC) and ejection fraction (FE), compared to simultaneous evaluation by invasive thermodilution (TD) in 39 patients with acute myocardial infarction in Killip class I-II (group I), and by cineventriculography (CVG) in 26 patients with chronic coronary artery disease in NYHA class I-II (group II). In order to define the reproducibility of TEB values, in the latter patients, the above mentioned parameters were evaluated 6 times more, running the first evaluation. The statistical analysis was performed by the linear regression test and the Student's "t" test and by the test of variance for the reproducibility evaluation. Results (mean +/- SD) were as follows: group I: TEB-IC 2.89 +/- 0.63; TD-IC 2.83 +/- 0.56 (1/min/m2); r = 0.68; p < 0.01. Group II: TEB-IC 2.88 +/- 0.71; CVG-IC 3.48 +/- 0.66; r = 0.77; p < 0.001; TEB-FE 57.7 +/- 6.8%; CVG-FE 58.1 +/- 13.7%; r = 0.40; p = ns. Results of the reproducibility referred to the 6 measurements (mean +/- SD) were the follows: TEB-IC (1/min/m2) (1) 2.83 +/- 0.76; (2) 2.85 +/- 0.73; (3) 2.8 +/- 0.79; (4) 2.83 +/- 0.71; (5) 2.87 +/- 0.81; (6) 2.88 +/- 0.8, p = ns, the variability was assesses within +/- 9.3%. TEB-FE (%): (1) 56.7 +/- 6.2; (2) 55.8 +/- 5; (3) 57.1 +/- 5.5; (4) 56.1 +/- 6.1; (5) 55.4 +/- 5.8; (6) 57.3 +/- 6.3, p = ns; the variability was assessed within +/- 9.1%. The analysis of the results showed a good correlation in the IC detection among TEB and the compared techniques, conversely TEB evaluation of FE appear of poor values in this kind of patients. Relatively to the results of the reproducibility this unquestionable characteristic of TEB was demonstrated.

摘要

相似文献

7
Measurement of cardiac output with non-invasive Aesculon impedance versus thermodilution.采用无创Aesculon阻抗法与热稀释法测量心输出量。
Clin Physiol Funct Imaging. 2011 Jan;31(1):39-47. doi: 10.1111/j.1475-097X.2010.00977.x. Epub 2010 Sep 23.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验