Suppr超能文献

用于计量吸入器喷雾粒径测量的双撞击器评估。

Assessment of the twin impinger for size measurement of metered-dose inhaler sprays.

作者信息

Miller N C, Marple V A, Schultz R K, Poon W S

机构信息

3M Pharmaceuticals Division, St. Paul, Minnesota 55144.

出版信息

Pharm Res. 1992 Sep;9(9):1123-7. doi: 10.1023/a:1015835301929.

Abstract

The calibration of the first stage of the twin-stage impinger, an instrument proposed for use in measuring the spray size from metered-dose inhalers, was performed with monodisperse aerosols by a standard technique for cascade impactors. The mean cut point was found to be not particularly sensitive to operating variables which may be expected to occur in practice. The cut point was close to that reported previously, although the collection efficiency curve was found to be slightly sharper. Calculations are reported on the expected results of measurements on aerosols in a two-stage instrument with an idealized perfect collection efficiency curve as well as the curve measured for the twin impinger. These results indicate that important characteristics of spray size distribution cannot be distinguished with an ideal two-stage instrument; the twin impinger is less capable than an ideal instrument.

摘要

双级冲击器的第一级校准是使用标准的级联冲击器技术,用单分散气溶胶对一种用于测量定量吸入器喷雾粒径的仪器进行的。发现平均切割点对实际中可能出现的操作变量不太敏感。切割点与先前报道的接近,尽管发现收集效率曲线略陡。报告了关于在具有理想化完美收集效率曲线的两级仪器以及双级冲击器测量曲线的情况下,对气溶胶测量预期结果的计算。这些结果表明,喷雾粒径分布的重要特征无法用理想的两级仪器区分;双级冲击器的能力不如理想仪器。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验