骨膜下吸收:手部X线片全帧图像压缩对诊断准确性的影响。

Subperiosteal resorption: effect of full-frame image compression of hand radiographs on diagnostic accuracy.

作者信息

Sayre J W, Ho B K, Boechat M I, Hall T R, Huang H K

机构信息

Department of Radiological Sciences, University of California, Los Angeles 90024-1721.

出版信息

Radiology. 1992 Nov;185(2):599-603. doi: 10.1148/radiology.185.2.1410380.

Abstract

Image compression is essential to handle a large volume of digital images, including computed tomographic, magnetic resonance, computed radiographic, and digitized images in a digital radiology operation. Developed during the past few years, full-frame bit allocation performed with the cosine transform technique has been proved to be an excellent irreversible image compression method. This article describes the effect, on the accuracy of diagnosis of subperiosteal resorption, of using the hardware compression module to produce hand radiographs. Receiver operating characteristic analysis of the interpretation of 71 radiographs by five observers demonstrated that there is no statistically significant difference in diagnostic accuracy between the original radiographs and compressed and reconstructed images obtained with a compression ratio as high as 20:1.

摘要

图像压缩对于处理大量数字图像至关重要,这些图像包括计算机断层扫描、磁共振、计算机射线摄影以及数字放射学操作中的数字化图像。在过去几年中开发的,采用余弦变换技术进行的全帧比特分配,已被证明是一种出色的不可逆图像压缩方法。本文描述了使用硬件压缩模块生成手部X光片对骨膜下骨质吸收诊断准确性的影响。对五名观察者解读的71张X光片进行的受试者操作特征分析表明,原始X光片与压缩比高达20:1的压缩和重建图像之间的诊断准确性没有统计学上的显著差异。