The effect of section alignment and distance in volumetric determination of myocardial mass by means of cardiac magnetic resonance imaging (MRI) was studied in 19 isolated pig hearts. Double angulated views taken through the short axis of the heart with a slice thickness of 10 mm and slices at a distance of 0 mm, 1 mm, 3 mm and 5 mm produced accurate left ventricular measurements with cardiac MRI. Measurements in the long axis of the heart resulted in a marked underestimation of myocardial mass; depending on slice intervals, there was an error between 13% and 20%. In vivo measurements in 35 subjects confirmed the effect of section alignment. Measurements in the long axis of the heart produced results of 139.9 +/- 31.7 g, significantly less than measurements in the short axis of the heart of 157.8 +/- 32.3 g.
通过心脏磁共振成像(MRI)测定心肌质量时,研究了19个离体猪心脏的切面方向和间距的影响。通过心脏短轴获取双角度视图,切片厚度为10毫米,切片间距为0毫米、1毫米、3毫米和5毫米,利用心脏MRI可对左心室进行准确测量。沿心脏长轴测量会导致心肌质量明显低估;根据切片间隔不同,误差在13%至20%之间。对35名受试者进行的活体测量证实了切面方向的影响。沿心脏长轴测量结果为139.9±31.7克,显著低于沿心脏短轴测量的157.8±32.3克。