Nitzsché R, Girodo S, Limousin M, Cazeau S
ELA Médical, Montrouge, France.
Pacing Clin Electrophysiol. 1992 Nov;15(11 Pt 2):1851-7. doi: 10.1111/j.1540-8159.1992.tb02981.x.
The methods used for preventing endless-loop tachycardias (ELTs) most often consist of initiating a long postventricular atrial refractory period (PVARP) with the sensing of every event likely to induce ELTs, such as sensed premature ventricular contractions (PVCs). A new fallback function may be useful to prevent the initiation of ELTs. A window of atrial rate acceleration detection (WARAD) is initiated with the sensing of every sinus event and equals 75% of the preceding PP interval. If an atrial event is sensed during this period, as are premature atrial contractions (PACs), no atrioventricular (AV) delay is initiated, but an atrial pulse output is delivered and a subsequent 31-msec AV delay is started. Theoretically retrograde P waves are premature compared to sinus rhythm. They are therefore detected as PACs, and do not initiate AV delay, thus prohibiting the induction of ELTs. This function was tested in six patients, using external or implanted Chorus 2 pacemakers. Short PVARP (203 msec) and high atrial sensibility were programmed. Retrograde conduction was induced either by inefficient atrial pacing or a long programmed AV delay. Two different dual chamber settings were tested: dual chamber pacing with the fallback function On or Off. In every situation, the function proved effective in preventing ELTs: the number of tachycardia episodes went from 124 with the function programmed Off to 5 with the function programmed On for comparable durations. More than 75 ELTs effectively prevented by fallback have been recorded.
预防无休止环形心动过速(ELTs)的方法通常包括通过感知每一个可能诱发ELTs的事件(如感知到的室性早搏(PVCs))来启动较长的心室后心房不应期(PVARP)。一种新的备用功能可能有助于预防ELTs的发生。心房率加速检测窗口(WARAD)在感知到每一个窦性事件时启动,其等于前一个PP间期的75%。如果在此期间感知到心房事件,如房性早搏(PACs),则不启动房室(AV)延迟,但会发放心房脉冲输出,并启动随后31毫秒的AV延迟。理论上,逆行P波相对于窦性心律来说是提前的。因此,它们被检测为PACs,并且不启动AV延迟,从而防止ELTs的诱发。使用外部或植入式Chorus 2起搏器对6名患者进行了该功能测试。设置了较短的PVARP(203毫秒)和较高的心房感知度。通过无效的心房起搏或较长的程控AV延迟来诱发逆行传导。测试了两种不同的双腔设置:备用功能开启或关闭的双腔起搏。在每种情况下,该功能都被证明能有效预防ELTs:在可比的持续时间内,心动过速发作次数从备用功能关闭时的124次降至备用功能开启时的5次。已记录到超过75次由备用功能有效预防的ELTs。