Brimacombe Joseph, Keller Christian
*Department of Anaesthesia and Intensive Care, James Cook University, Cairns Base Hospital, Australia; and †Department of Anaesthesia and Intensive Care Medicine, Leopold-Franzens University, Innsbruck, Austria.
Anesth Analg. 2003 Oct;97(4):1192-1194. doi: 10.1213/01.ANE.0000078818.64971.44.
The primary function of the drain tube of the ProSeal laryngeal mask airway (PLMA) is to prevent aspiration of regurgitated stomach contents. We report a case of gastric aspiration with the PLMA during a laparoscopic cholecystectomy in a 64-yr-old healthy man secondary to an unidentified foldover malposition. It is imperative that the position and patency of the drain tube be verified in all patients with the PLMA.
We report a case of gastric aspiration with the ProSeal laryngeal mask during a laparoscopic cholecystectomy secondary to an unidentified foldover malposition. It is imperative that the position and patency of the drain tube be verified in all patients with the ProSeal laryngeal mask.
食管引流型喉罩气道(PLMA)的引流管主要功能是防止胃内容物反流误吸。我们报告一例64岁健康男性在腹腔镜胆囊切除术期间使用PLMA时发生胃内容物误吸的病例,原因是未识别出的折叠错位。对所有使用PLMA的患者,必须确认引流管的位置及通畅情况。
我们报告一例在腹腔镜胆囊切除术期间因未识别出的折叠错位导致使用ProSeal喉罩发生胃内容物误吸的病例。对所有使用ProSeal喉罩的患者,必须确认引流管的位置及通畅情况。