乳腺钼靶筛查中的计算机辅助检测(CAD):商用CAD系统检测结构扭曲的敏感性

Computer-aided detection (CAD) in screening mammography: sensitivity of commercial CAD systems for detecting architectural distortion.

作者信息

Baker Jay A, Rosen Eric L, Lo Joseph Y, Gimenez Edgardo I, Walsh Ruth, Soo Mary Scott

机构信息

Department of Radiology, Duke University Medical Center, Box 3808, Durham, NC 27710, USA.

出版信息

AJR Am J Roentgenol. 2003 Oct;181(4):1083-8. doi: 10.2214/ajr.181.4.1811083.

Abstract

OBJECTIVE

Computer-aided detection (CAD) algorithms have successfully revealed breast masses and microcalcifications on screening mammography. The purpose of our study was to evaluate the sensitivity of commercially available CAD systems for revealing architectural distortion, the third most common appearance of breast cancer.

MATERIALS AND METHODS

Two commercially available CAD systems were used to evaluate screening mammograms obtained in 43 patients with 45 mammographically detected regions of architectural distortion. For each CAD system, we determined the sensitivity for revealing architectural distortion on at least one image of the two-view mammographic examination (case sensitivity) and for each individual mammogram (image sensitivity). Surgical biopsy results were available for each case of architectural distortion.

RESULTS

Architectural distortion was deemed present and actionable by a panel of expert breast imagers in 80 views of the 45 cases. One CAD system detected distortion in 22 of 45 cases of distortion (case sensitivity, 49%) and in 30 of 80 mammograms (image sensitivity, 38%); it displayed 0.7 false-positive marks per image. Another CAD system identified distortion in 15 of 45 cases (case sensitivity, 33%) and 17 of 80 mammograms (image sensitivity, 21%); it displayed 1.27 false-positive marks per image. Sensitivity for malignancy-caused distortion was similar to or lower than sensitivity for all causes of distortion.

CONCLUSION

Fewer than one half of the cases of architectural distortion were detected by the two most widely available CAD systems used for interpretations of screening mammograms. Considerable improvement in the sensitivity of CAD systems is needed for detecting this type of lesion. Practicing breast imagers who use CAD systems should remain vigilant for architectural distortion.

摘要

目的

计算机辅助检测(CAD)算法已成功在乳腺钼靶筛查中发现乳腺肿块和微钙化。我们研究的目的是评估市售CAD系统检测乳腺结构扭曲的敏感性,乳腺结构扭曲是乳腺癌的第三大常见表现。

材料与方法

使用两种市售CAD系统评估43例患者的乳腺钼靶筛查图像,这些图像中有45个在钼靶上检测到的乳腺结构扭曲区域。对于每个CAD系统,我们确定其在双视图乳腺钼靶检查的至少一张图像上检测乳腺结构扭曲的敏感性(病例敏感性)以及每张乳腺钼靶图像的敏感性(图像敏感性)。每例乳腺结构扭曲病例均有手术活检结果。

结果

在45例病例的80张图像中,一组乳腺影像专家认为存在乳腺结构扭曲且需要采取行动。一种CAD系统在45例扭曲病例中的22例中检测到扭曲(病例敏感性为49%),在80张钼靶图像中的30张中检测到扭曲(图像敏感性为38%);每张图像显示0.7个假阳性标记。另一种CAD系统在45例病例中的15例中识别出扭曲(病例敏感性为33%),在80张钼靶图像中的17张中识别出扭曲(图像敏感性为21%);每张图像显示1.27个假阳性标记。恶性肿瘤导致的扭曲的敏感性与所有原因导致的扭曲的敏感性相似或更低。

结论

用于解读乳腺钼靶筛查的两种最常用CAD系统检测到的乳腺结构扭曲病例不到一半。需要大幅提高CAD系统检测此类病变的敏感性。使用CAD系统的乳腺影像科医生在检查时应留意乳腺结构扭曲。