Ketelslegers Etienne, Van Beers Bernard E
Diagnostic Radiology Unit, Cliniques Universitaires St-Luc, Université Catholique de Louvain, Avenue Hippocrate 10, 1200, Brussels, Belgium.
Eur Radiol. 2006 Jan;16(1):161-5. doi: 10.1007/s00330-005-2813-y. Epub 2005 Jun 16.
The aim of this study was to assess the effect of reconstructed slice thickness on the detection and characterization of human urinary calculi on a multidetector helical CT scanner. Nineteen human urinary calculi of various chemical composition measuring 1.0-3.7 mm were embedded into agar in a chamber of a nylon body phantom. The phantom was imaged with a four detector-row CT scanner. The number of detected calculi increased as the reconstructed slice thickness decreased. Measured diameters and density of the visible calculi decreased as the slice thickness increased. The results of the present study support the use of thin reconstructed slices to detect and characterize urinary calculi.