Suppr超能文献

采用替代袖带配置减少伪影。

Artefact reduction with alternative cuff configurations.

作者信息

Andreasen Lotte N S, Struijk Johannes J

机构信息

Center for Sensory Motor Interaction, Aalborg University, Denmark.

出版信息

IEEE Trans Biomed Eng. 2003 Oct;50(10):1160-6. doi: 10.1109/TBME.2003.817633.

Abstract

In nerve cuff electrode recordings of neural signals, the pick-up of interfering signals can be reduced by choosing appropriate cuff configurations. In the traditionally used tripolar configuration, short circuiting of the end electrodes is expected to reduce the field inside the cuff from interfering signals. A model study suggests that moving the end electrodes toward the center of the cuff reduces the pick-up of interfering signals. In this paper, these properties are studied in more detail using a rabbit model. In addition, a new cuff configuration is suggested, which has an additional set of short circuited end electrodes. The total improvement of signal-to-noise ratio in the new configuration as compared with the traditionally used tripolar configuration was 73% for muscle signals and 127% for the stimulus pulse.

摘要

在神经信号的神经袖套电极记录中,通过选择合适的袖套配置可以减少干扰信号的拾取。在传统使用的三极配置中,预期末端电极的短路会减少袖套内来自干扰信号的场。一项模型研究表明,将末端电极向袖套中心移动可减少干扰信号的拾取。在本文中,使用兔模型对这些特性进行了更详细的研究。此外,还提出了一种新的袖套配置,它有一组额外的短路末端电极。与传统使用的三极配置相比,新配置中肌肉信号的信噪比总体提高了73%,刺激脉冲的信噪比提高了127%。