Pan Yingtian, Li Zhigang, Xie Tuqiang, Chu Constance R
State University of New York, Stony Brook, Department of Biomedical Engineering, Stony Brook, New York 11794-8181, USA.
J Biomed Opt. 2003 Oct;8(4):648-54. doi: 10.1117/1.1609201.
We describe a novel hand-held polarization optical coherence tomographic (OCT) probe that can be inserted into mammalian joints to permit real-time cross-sectional imaging of articular cartilage. The transverse and axial resolutions of the arthroscopic OCT device are roughly 17 and 10 microm, respectively. Two-dimensional cross-sectional images of cartilage tissue with 500 x 1000 pixels covering an area 6 mm in length and 2.8 mm in depth can be acquired at nearly five frames/s and with over 100 dB of dynamic range. Design of an OCT as a hand-held device capable of providing such an optical biopsy of articular cartilage allows eventual in vivo detection of microstructural changes in articular cartilage that are not apparent using conventional arthroscopic cameras. The OCT probe can be easily incorporated in a conventional arthroscope for cartilage site guidance. The optical arrangement in the OCT scope minimizes specular back-reflection of the probe end face and absorption of body fluid in the path and ensures in-focus OCT imaging when it is in contact with the cartilage specimen to be examined. Successful application of in vivo arthroscopy to porcine articular cartilage demonstrates sufficient resolution and practicality for use in human joints.
我们描述了一种新型手持式偏振光学相干断层扫描(OCT)探头,该探头可插入哺乳动物关节,以实现关节软骨的实时横截面成像。关节镜OCT设备的横向分辨率和轴向分辨率分别约为17微米和10微米。可以以近每秒五帧的速度采集覆盖长度6毫米、深度2.8毫米区域的500×1000像素的软骨组织二维横截面图像,动态范围超过100分贝。将OCT设计为能够对关节软骨进行这种光学活检的手持式设备,最终可在体内检测到使用传统关节镜相机无法发现的关节软骨微观结构变化。OCT探头可以很容易地集成到传统关节镜中,用于软骨部位引导。OCT镜中的光学装置可将探头端面的镜面背反射和光路中体液的吸收降至最低,并确保在与待检查的软骨标本接触时进行聚焦OCT成像。在猪关节软骨上成功应用体内关节镜检查证明了其在人体关节中使用具有足够的分辨率和实用性。