Nordin S, Zulkifli O, Faisham W I
Department of Orthopaedics, School of Medical Science, Hospital Universiti Sains Malaysia, 16150 Kubang Kerian, Kelantan.
Med J Malaysia. 2001 Dec;56 Suppl D:12-7.
We studied 60 intertrochanteric fractures of the femur fixed with Dynamic Hip Screw (DHS). There were 10 cases (16.7%) with cutting-out of device through femoral head and neck. Stable fracture pattern, postero-inferior and central position of screw in the femoral neck and head produced high percentage of good result, whereas anterior or superior position of screw produced higher incidence of cut-out. We found osteoporosis and distance of screw tip to subchondral bone to have no influence on the final outcome.
我们研究了60例采用动力髋螺钉(DHS)固定的股骨粗隆间骨折。其中有10例(16.7%)出现内固定装置穿出股骨头和颈部的情况。稳定的骨折类型、螺钉位于股骨颈和头部的后下及中央位置,取得良好结果的比例较高,而螺钉位于前方或上方位置时,穿出的发生率较高。我们发现骨质疏松以及螺钉尖端至软骨下骨的距离对最终结果并无影响。