Poitout V, Moatti-Sirat D, Reach G
INSERM U341, Service de Diabétologie, Hôtel-Dieu, Paris, France.
Biosens Bioelectron. 1992;7(8):587-92. doi: 10.1016/0956-5663(92)85010-8.
The feasibility of calibrating a glucose sensor by using a wearable glucose meter for blood glucose determination and moderate variations of blood glucose concentration was assessed. Six miniaturized glucose sensors were implanted in the subcutaneous tissue of conscious dogs, and the parameters used for the in vivo calibration of the sensor (sensitivity coefficient and extrapolated current in the absence of glucose) were determined from values of blood glucose and sensor response obtained during glucose infusion. (1) Venous plasma glucose level and venous total blood glucose level were measured simultaneously on the same sample, using a Beckman analyser and a Glucometer II, respectively. The regression between plasma glucose (x) and whole blood glucose (y) was y = 1.12x-0.08 mM (n = 114 values, r = 0.96, p = 0.0001). The error grid analysis indicated that the use of a Glucometer II for blood glucose determination was appropriate in dogs. (2) The in vivo sensitivity coefficients were 0.57 +/- 0.11 nA mM-1 when determined from plasma glucose, and 0.51 +/- 0.07 nA mM-1 when determined from whole blood glucose (t = 1.53, p = 0.18, n.s.). The background currents were 0.88 +/- 0.57 nA when determined from plasma glucose, and 0.63 +/- 0.77 nA when determined from whole blood glucose (t = 0.82, p = 0.45, n.s.). (3) The regression equation of the estimation of the subcutaneous glucose level obtained from the two methods was y = 1.04x + 0.56 mM (n = 171 values, r = 0.98, p = 0.0001).(ABSTRACT TRUNCATED AT 250 WORDS)
评估了使用可穿戴血糖仪校准葡萄糖传感器以测定血糖及血糖浓度适度变化的可行性。将六个小型化葡萄糖传感器植入清醒犬的皮下组织,根据葡萄糖输注期间获得的血糖值和传感器响应确定用于传感器体内校准的参数(灵敏度系数和无葡萄糖时的外推电流)。(1)分别使用贝克曼分析仪和血糖仪II在同一样本上同时测量静脉血浆葡萄糖水平和静脉全血葡萄糖水平。血浆葡萄糖(x)与全血葡萄糖(y)之间的回归方程为y = 1.12x - 0.08 mM(n = 114个值,r = 0.96,p = 0.0001)。误差网格分析表明,在犬中使用血糖仪II测定血糖是合适的。(2)根据血浆葡萄糖测定的体内灵敏度系数为0.57±0.11 nA mM-1,根据全血葡萄糖测定的为0.51±0.07 nA mM-1(t = 1.53,p = 0.18,无显著性差异)。根据血浆葡萄糖测定的背景电流为0.88±0.57 nA,根据全血葡萄糖测定的为0.63±0.77 nA(t = 0.82,p = 0.45,无显著性差异)。(3)两种方法获得的皮下葡萄糖水平估计回归方程为y = 1.04x + 0.56 mM(n = 171个值,r = 0.98,p = 0.0001)。(摘要截断于250字)