水凝胶隐形眼镜折射率自动测量方法的评估

The assessment of automated measures of hydrogel contact lens refractive index.

作者信息

Nichols Jason J, Berntsen David A

机构信息

College of Optometry, The Ohio State University, 320 West 10th Avenue, PO Box 182342, Columbus, OH 43218-2342, USA.

出版信息

Ophthalmic Physiol Opt. 2003 Nov;23(6):517-25. doi: 10.1046/j.1475-1313.2003.00147.x.

Abstract

PURPOSE

To investigate within- and between-operator reliability and validity of refractive index measures of hydrogel lenses.

METHODS

Twenty-three lenses of various nominal water contents were examined by two operators on two occasions separated by 1 h. An automated refractometer (CLR 12-70, Index Instruments, Cambridge, UK) was used for refractive index measures. Lenses were presented in a random order to each operator by a third party, and operators were masked to any potential lens identifiers. Intraclass correlation coefficients (ICC), 95% limits of agreement (LoA), and paired t-tests were used to characterize the within- and between-operator reliability and validity of lens refractive index measures.

RESULTS

The mean (+/-S.D.) difference between sessions was 0.0001 +/- 0.0011, which did not significantly differ from zero (p = 0.64). Within-operator reliability was excellent (ICC = 0.9990, 95% LoA = -0.0021 to 0.0024). The mean (+/-S.D.) difference between-operators was 0.0002 +/- 0.0010, which was not significantly different from zero (p = 0.54). Between-operator reliability was also excellent (ICC = 0.9984, 95% LoA = -0.0026 to 0.0029). The mean difference between the nominally reported refractive index and our refractive index measures was -0.0009 +/- 0.082, which did not differ statistically from zero (p = 0.52); the 95% LoA for this difference was -0.0152 to 0.0169.

CONCLUSION

There is good reliability within- and between-operators in measuring the refractive index of hydrogel lenses using the CLR 12-70 automated refractometer.

摘要

目的

研究水凝胶镜片折射率测量在操作者内部及操作者之间的可靠性和有效性。

方法

两名操作者在两次测量中对23片不同标称含水量的镜片进行检测,两次测量间隔1小时。使用自动折射仪(CLR 12 - 70,Index Instruments,英国剑桥)测量折射率。第三方以随机顺序向每位操作者提供镜片,操作者对任何潜在的镜片标识不知情。组内相关系数(ICC)、95%一致性界限(LoA)和配对t检验用于表征镜片折射率测量在操作者内部及操作者之间的可靠性和有效性。

结果

两次测量之间的平均(±标准差)差值为0.0001±0.0011,与零无显著差异(p = 0.64)。操作者内部可靠性极佳(ICC = 0.9990,95% LoA = - 0.0021至0.0024)。操作者之间的平均(±标准差)差值为0.0002±0.0010,与零无显著差异(p = 0.54)。操作者之间的可靠性也极佳(ICC = 0.9984,95% LoA = - 0.0026至0.0029)。标称报告的折射率与我们测量的折射率之间的平均差值为 - 0.0009±0.082,与零无统计学差异(p = 0.52);该差值的95% LoA为 - 0.0152至0.0169。

结论

使用CLR 12 - 70自动折射仪测量水凝胶镜片的折射率时,在操作者内部及操作者之间具有良好的可靠性。