Cox Guy, Sheppard Colin J R
Instituto Gulbenkian de Ciencia, Oeiras, Portugal.
Microsc Res Tech. 2004 Jan 1;63(1):18-22. doi: 10.1002/jemt.10423.
Calculated and measured resolution figures are presented for confocal microscopes with different pinhole sizes and for nonlinear (2-photon and second harmonic) microscopes. A modest degree of super-resolution is predicted for a confocal microscope but in practice this is not achievable and confocal fluorescence gives little resolution improvement over widefield. However, practical non-linear microscopes do approach their theoretical resolution and therefore show no resolution disadvantage relative to confocal microscopes in spite of the longer excitation wavelength.