[Study of ultrasound attenuation during HIFU propagation in ox liver].

作者信息

Li Faqi, Bai Jin, Wang Zhibiao, Ma Ping, Du Yonghong, Wu Feng, Feng Ruo

机构信息

Institute of Ultrasonic Engineering in Medicine, Chongqing University of Medical Science, Chongqing 400016.

出版信息

Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2003 Dec;20(4):675-8.

Abstract

Ultrasound attenuation has been measured during HIFU propagation in fresh ox liver at different frequency using a radiation force method. The acoustic radiation force was measured before and after the prepared ox liver of various thicknesses (20 mm, 40 mm, 60 mm, 80 mm) which were put into degassed water and exposed to different transducer surface output acoustic powers at a room temperature (20 degrees C) with therapeutic transducers of various frequencies, and then the ultrasound attenuation coefficient was calculated. With the use of therapeutic transducer 4, the focus of the ultrasonic beam was set to be 20 mm, 40 mm or 60 mm deep into the tissue surface using B-mode ultrasound guidance. A single exposure was performed with focal intensity ISATA = 22.0 x 10(3) W/cm2(ISATA, in degassed water) and exposure time 5 sec. The sample was cut after treatment to measure the volume of coagulative necrosis. For a specific therapeutic transducer, the radiation force ratio in liver of constant thickness is independent of the transducer surface sound intensity and the area of the ultrasound window at the sample front face. The radiation force ratio and the volume of coagulative necrosis induced by HIFU are plotted as the functions of sample thickness (or exposed depth), in which it can be seen that these two parameters have the same exponential dependence on sample thickness. The ultrasound attenuation coefficient alpha in ox liver is shown to be frequency f dependent, it almost linearly increases with frequency t. This work shows that such a study is feasible, and offers experimental data that will be useful for future HIFU dosage studies.

摘要

使用辐射力方法,在不同频率下测量了高强度聚焦超声(HIFU)在新鲜牛肝中传播时的超声衰减。在室温(20摄氏度)下,将不同厚度(20毫米、40毫米、60毫米、80毫米)的制备好的牛肝放入脱气水中,使用不同频率的治疗换能器,在不同的换能器表面输出声功率下,测量声辐射力,然后计算超声衰减系数。使用治疗换能器4,在B型超声引导下,将超声束的焦点设置在组织表面以下20毫米、40毫米或60毫米处。以焦强ISATA = 22.0×10³W/cm²(在脱气水中的ISATA)和暴露时间5秒进行单次暴露。治疗后切割样本以测量凝固性坏死的体积。对于特定的治疗换能器,恒定厚度肝脏中的辐射力比与换能器表面声强以及样本正面超声窗口的面积无关。将HIFU诱导的辐射力比和凝固性坏死体积绘制为样本厚度(或暴露深度)的函数,从中可以看出这两个参数对样本厚度具有相同的指数依赖性。牛肝中的超声衰减系数α显示为与频率f相关,它几乎随频率f线性增加。这项工作表明这样的研究是可行的,并提供了对未来HIFU剂量研究有用的实验数据。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验