Enlargement of thermal ablation zone by the combination of ethanol injection and radiofrequency ablation in excised bovine liver.

作者信息

Watanabe Seishiro, Kurokohchi Kazutaka, Masaki Tsutomu, Miyauchi Yoshiaki, Funaki Toshiharu, Inoue Hideyuki, Himoto Takashi, Kita Yuko, Uchida Naohito, Touge Tetsuo, Tatsukawa Toshiaki, Kuriyama Shigeki

机构信息

Third Department of Internal Medicine, Kagawa University School of Medicine, 1750-1 Ikenobe, Miki-cho, Kita-gun, Kagawa 761-0793, Japan.

出版信息

Int J Oncol. 2004 Feb;24(2):279-84.

Abstract

The efficacy of radiofrequency ablation (RFA) and RFA with concurrent ethanol injection (EI-RFA) was compared. RFA (3-cm-electrode) was applied to bovine liver using three types of RFA equipment; Radionics, RITA and Radio Therapeutics Corporation (RTC). For EI-RFA, 5 ml of 99.5% ethanol was injected around the electrode. A total number of 40 RFA and EI-RFA treatments were performed. We compared RFA with EI-RFA by examining the size, shape of ablation zone, treatment time, power, and needle tip temperature. Liver specimens were examined for pathological changes. EI-RFA produced a larger zone of ablation than RFA alone using Radionics and RITA (Radionics, 35.3+/-7.4 cm(3) vs 23.2+/-7.7 cm(3), p<0.05; RITA, 30.7+/-10.3 cm(3) vs 19.7+/-4.7 cm(3), p<0.05), corresponding to shortest diameters of coagulation zone (Radionics, 3.7+/-0.4 cm vs 3.0+/-0.4 cm, p<0.05; RITA, 3.8+/-0.4 cm vs 3.1+/-0.3 cm, p<0.01). However, a larger ablation zone was not seen with the RTC device. The ablated volume per energy and the ablated volume per current density administered were greater with EI-RFA than with RFA using Radionics (p<0.05). The shape of the ablated zone changed from ellipsoid to spherical with EI-RFA using Radionics. No pathological differences between RFA and EI-RFA samples were detected. For a given amount of energy and current administered, ethanol injection caused a better ablation effect, in terms of the size and shape of the ablated zone, than RFA with Radionics and RITA equipment.

摘要

比较了射频消融(RFA)与同时乙醇注射的射频消融(EI-RFA)的疗效。使用三种类型的RFA设备(Radionics、RITA和放射治疗公司(RTC))将RFA(3厘米电极)应用于牛肝脏。对于EI-RFA,在电极周围注射5毫升99.5%的乙醇。共进行了40次RFA和EI-RFA治疗。我们通过检查消融区的大小、形状、治疗时间、功率和针尖温度来比较RFA和EI-RFA。对肝脏标本进行病理变化检查。与单独使用Radionics和RITA的RFA相比,EI-RFA产生的消融区更大(Radionics,35.3±7.4立方厘米对23.2±7.7立方厘米,p<0.05;RITA,30.7±10.3立方厘米对19.7±4.7立方厘米,p<0.05),对应于凝固区的最短直径(Radionics,3.7±0.4厘米对3.0±0.4厘米,p<0.05;RITA,3.8±0.4厘米对3.1±0.3厘米,p<0.01)。然而,使用RTC设备未观察到更大的消融区。与使用Radionics的RFA相比,EI-RFA每能量和每电流密度的消融体积更大(p<0.05)。使用Radionics的EI-RFA使消融区的形状从椭圆形变为球形。未检测到RFA和EI-RFA样本之间的病理差异。对于给定的能量和电流剂量,就消融区的大小和形状而言,乙醇注射比使用Radionics和RITA设备的RFA产生更好的消融效果。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验