Dudziński Andrzej, Zawojska Izabela, Kinasz Renata
I Katedry i Kliniki Okulistyki Slaskiej Akademii Medycznej w Katowicach.
Klin Oczna. 2003;105(5):283-7.
Evaluation of CFF perimetry results in patients with suspected glaucoma.
Visual field was measured on Octopus 301 perimeter with option "flicker perimetry". 49 patients (98 eyes) with suspected glaucoma were tested. Standard W/W perimetry revealed glaucomatous visual field defect in one eye, but not in the second eye. Patients with cataract and diabetes were not taken into consideration.
Defects in CFF perimetry tests were found in 71.4% of cases, but in W/W perimetry test in 50.0% of cases.
Perimetry CFF is useful for early detection and follow up of glaucoma. Indices MD and LV may be used for those purposes.
评估疑似青光眼患者的闪光融合频率视野检查结果。
使用配备“闪光融合频率视野检查”选项的Octopus 301视野计测量视野。对49例(98只眼)疑似青光眼患者进行了检测。标准的W/W视野检查在一只眼中发现了青光眼性视野缺损,但另一只眼中未发现。未考虑患有白内障和糖尿病的患者。
闪光融合频率视野检查中71.4%的病例发现有缺损,而W/W视野检查中50.0%的病例发现有缺损。
闪光融合频率视野检查对青光眼的早期检测和随访有用。平均缺损(MD)和丢失方差(LV)指标可用于这些目的。