Hyde R J, Lohse C L, Testoni F J
Am J Physiol. 1978 May;234(5):F452-4. doi: 10.1152/ajprenal.1978.234.5.F452.
The small size of a new thermistor flowmeter facilitated implantation in ureters in anesthetized dogs. It consisted of a self-heating thermistor bead which was suspended centrally within an open catheter and dissipated heat as a function of fluid velocity. A continuous record of instantaneous flow rate was obtained, since the flowmeter catheter maintained a constant cross-sectional area around the thermistor bead. A continuous record of volume output was obtained simultaneously with an additional measurement technique. Spurt volumes of urine delivered in concert with ureter peristalsis were directly correlated to peak flow rates delineated by the flowmeter.
一种新型热敏电阻流量计体积小巧,便于植入麻醉犬的输尿管。它由一个自热式热敏电阻珠组成,该电阻珠悬挂在开放导管的中心位置,并根据流体速度散热。由于流量计导管在热敏电阻珠周围保持恒定的横截面积,因此可以获得瞬时流速的连续记录。通过另一种测量技术同时获得了体积输出的连续记录。与输尿管蠕动同步排出的尿液喷射量与流量计描绘的峰值流速直接相关。