An isothermal flowmeter with improved frequency response for measuring tissue blood flow.

作者信息

Olshausen K, Gross R, Kirchheim H

出版信息

Pflugers Arch. 1976 Nov 30;367(1):97-102. doi: 10.1007/BF00583661.

Abstract

An isothermal flowmeter with improved frequency response for measuring tissue blood flow was developed using thermistors. Direct heating of the thermistors allows a simple construction of small (0.5 mm outer diameter) capillary probes which do not require any additional heating coil. The changes of a feedback current necessary to keep the thermistor at a constant increment above tissue temperature indicate tissue blood flow; a second thermistor compensates variations of tissue temperature. The dynamic performance of the device shows a low-pass characteristic with a cut-off frequency higher than 5 Hz. For low flow rates the output signal was found to be proportional to the flow; for higher flow rates a linearization was necessary. Since tissue temperature can be recorded continuously, intermittent quantitative in-vivo calibration seems possibly by evaluation of "heater off" curves in the perfused and non-perfused tissue. As the flowmeter is insensitive to tissue temperature, it can be used for long-term recordings.