A compact centrifugal pump for cardiopulmonary bypass.

作者信息

Sasaki T, Jikuya T, Aizawa T, Shiono M, Sakuma I, Takatani S, Glueck J, Noon G P, Nosé Y, DeBakey M E

机构信息

Baylor College of Medicine, Department of Surgery, Houston, Texas 77030.

出版信息

Artif Organs. 1992 Dec;16(6):592-8. doi: 10.1111/j.1525-1594.1992.tb00557.x.

Abstract

A majority of the cardiopulmonary bypass (CPB) systems still utilize bulky roller pumps. A direct-drive small centrifugal pump intended for second-generation CPB pump has been developed. The pump has a 50 mm diameter impeller and provides a 6 L/min flow at 3,000 rpm against 300 mm Hg. A flexible drive shaft allows us to separate the pump head from the console resulting in easier manipulation. An in vitro study showed that the pump generated less hemolysis (index of hemolysis = 0.0011, comparable to the value for Bio-medicus BP-80). To improve blood flow around the shaft-seal region and to reduce thrombus formation around the shaft, six holes were drilled through the impeller. In biventricular bypass experiments using calves, our pump demonstrated excellent antithrombogenicity and durability for 48 h. And the compact and atraumatic centrifugal pump system showed excellent performance and easy manipulation under actual CPB conditions in animal.

摘要

大多数体外循环(CPB)系统仍在使用笨重的滚压泵。一种用于第二代CPB泵的直接驱动小型离心泵已被研发出来。该泵有一个直径50毫米的叶轮,在3000转/分钟时可提供6升/分钟的流量,能克服300毫米汞柱的压力。一根柔性驱动轴使我们能够将泵头与控制台分离,从而便于操作。一项体外研究表明,该泵产生的溶血较少(溶血指数 = 0.0011,与Bio-medicus BP - 80的值相当)。为了改善轴封区域周围的血流并减少轴周围的血栓形成,在叶轮上钻了六个孔。在使用小牛的双心室旁路实验中,我们的泵在48小时内表现出了出色的抗血栓形成能力和耐用性。并且这种紧凑且无创伤的离心泵系统在动物实际CPB条件下表现出了出色的性能和易于操作的特点。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验