Franke M, Goenechea S, Heinbach F
Institut für Rechtsmedizin, Universität Bonn.
Beitr Gerichtl Med. 1992;50:105-9.
In the last 2 years we had some problems to confirm the thin-layer results (amphetamines, opiates, benzodiazepines, barbiturates and cocaine) with the immunoassay system TDx/ADx. Between January 1989 and December 1990 we did 883 screenings in urine samples (pre- and postmortal) and had 157 discrepancies (20%). In regard of the high level of discrepancies, we had to test the compatibility of TDx and ADx, their detection limits and compared the results with a second method, thin-layer chromatography. Determining the correlation coefficients to prove our results statistically, we found that both systems, TDx and ADx, work reproducible.
在过去两年中,我们在使用免疫分析系统TDx/ADx确认薄层分析结果(安非他明、阿片类药物、苯二氮卓类、巴比妥类和可卡因)时遇到了一些问题。1989年1月至1990年12月期间,我们对尿液样本(生前和死后)进行了883次筛查,出现了157次差异(20%)。鉴于差异水平较高,我们必须测试TDx和ADx的兼容性、它们的检测限,并将结果与第二种方法薄层色谱法进行比较。为了从统计学上证明我们的结果而确定相关系数时,我们发现TDx和ADx这两种系统都具有可重复性。