Larsson P U, Wadell K M E, Jakobsson E J I, Burlin L U, Henriksson-Larsén K B
Sports Medicine Unit, University of Umeå, Sweden.
Int J Sports Med. 2004 Feb;25(2):115-23. doi: 10.1055/s-2004-819953.
The purpose of this study was to investigate the validity of the MetaMax II portable metabolic measurement system against the Douglas Bag technique. Nine recreationally active male subjects were included in a validation at 100 W, 10 well-trained male subjects at 200 W and 10 well-trained males at 250 W and at maximal exercise (volitional fatigue at a mean workload of 325 W). All testing was performed on an electronically braked bicycle at 60 rpm. At 100 W, the influence on MetaMax II measurements of adding a Douglas Bag breathing valve in series to the MetaMax II was investigated. The oxygen uptake was, for the MetaMax II, at 100 W mean 0.03 l x min (-1) higher (p < 0.01), at 200 W mean 0.02 l x min (-1) (n. s.) lower, at 250 W mean 0.04 l x min (-1) (n. s.) higher, and at 325 W mean 0.11 l x min (-1) (p < 0.05) higher. The carbon dioxide excretion was, for the MetaMax II, at 100 W mean 0.06 l x min (-1) (p < 0.01) lower, at 200 W mean 0.11 l x min (-1) (p < 0.05) lower, at 250 W mean 0.03 l x min (-1) (n. s.) lower, and at 325 W mean 0.16 l x min (-1) (p < 0.05) lower. The addition of a breathing valve in series to the MetaMax II resulted in lower breathing frequency, a higher ventilated tidal volume, and an affected gas measurement validation. In conclusion, the MetaMax II was found to be valid for metabolic gas measurements between 100 and at least 250 W.
本研究旨在探讨MetaMax II便携式代谢测量系统相对于道格拉斯袋技术的有效性。九名有休闲运动习惯的男性受试者参与了100瓦功率下的验证测试,十名训练有素的男性受试者参与了200瓦功率下的测试,另有十名训练有素的男性受试者参与了250瓦功率及最大运动强度(平均工作负荷为325瓦时的自愿疲劳)下的测试。所有测试均在电子刹车自行车上以60转/分钟的速度进行。在100瓦功率下,研究了在MetaMax II上串联一个道格拉斯袋呼吸阀对其测量结果的影响。对于MetaMax II,氧气摄取量在100瓦时平均高0.03升/分钟(p < 0.01),在200瓦时平均低0.02升/分钟(无显著差异),在250瓦时平均高0.04升/分钟(无显著差异),在325瓦时平均高0.11升/分钟(p < 0.05)。二氧化碳排出量对于MetaMax II,在100瓦时平均低0.06升/分钟(p < 0.01),在200瓦时平均低0.11升/分钟(p < 0.05),在250瓦时平均低0.03升/分钟(无显著差异),在325瓦时平均低0.16升/分钟(p < 0.05)。在MetaMax II上串联一个呼吸阀会导致呼吸频率降低、潮气量增加,并影响气体测量的验证。总之,发现MetaMax II在100至至少250瓦之间的代谢气体测量中是有效的。