Suppr超能文献

两种便携式呼吸气体分析仪的比较。

Comparison of 2 portable respiratory gas analysers.

机构信息

Laboratoire de Recherche Adaptations Physiologiques à l'Exercice et -Réadaptation à l'Effort Université de Picardie Jules Verne, AMIENS, France.

出版信息

Int J Sports Med. 2012 Sep;33(9):728-33. doi: 10.1055/s-0031-1301316. Epub 2012 May 4.

Abstract

The aim of this study was to compare respiratory measures taken simultaneously using the Cosmed K4b(2) and Cortex Metamax II portable metabolic systems. 10 trained male cyclists performed a graded exercise cycle test to exhaustion (40w.3 min (- 1)) under standardized conditions. The measured respiratory variables were significantly correlated between both devices: r=0.97 and 0.98 (n=10, p<0.01) for oxygen uptake ( V˙O(2)) and ventilation (VE), respectively. Further Bland and Altman plots revealed a good level of agreement for measures of V˙O(2) expressed in mL.min (- 1) [- 670 to 486] (mean bias of - 91.7, i. e., - 3.1%) or in mL.min (- 1).kg (- 1) [- 7.3 to 9.0] (mean bias of 0.85, i. e., 2.3%) and V(E) [- 23.1 to 18.2] (L.min (- 1), mean bias of - 2.4, i. e. - 4.1%). However, poor agreement was found for measures of carbon dioxide (VCO(2), mL.min (- 1)) [- 280 to 1 394] (mean bias of 671 i.e., 20.3%). VCO(2) at maximal exercise intensity was also significantly (p<0.01) greater in the Cortex compared to the Cosmed system. The higher measured CO(2) concentrations led to significantly (p<0.01) higher calculated respiratory exchange ratio (RER) values with the Cortex device. In conclusion, there was satisfactory agreement between the Cosmed K4b(2) and Cortex Metamax II systems for most respiratory measures; however there was a poor level of agreement between VCO(2) and calculated RER measurements between the 2 systems.

摘要

本研究旨在比较同时使用 Cosmed K4b(2) 和 Cortex Metamax II 便携式代谢系统进行的呼吸测量。10 名经过训练的男性自行车运动员在标准条件下进行了分级运动自行车测试直至力竭(40w.3 min(-1))。测量的呼吸变量在两个设备之间具有显著相关性:r=0.97 和 0.98(n=10,p<0.01),分别为氧气摄取量(V˙O(2))和通气量(VE)。进一步的 Bland 和 Altman 图显示,V˙O(2)的测量值在以 mL.min(-1)表示时([-670 至 486],平均偏差为-91.7,即-3.1%)或在以 mL.min(-1).kg(-1)表示时([-7.3 至 9.0],平均偏差为 0.85,即 2.3%)和 VE([-23.1 至 18.2],L.min(-1),平均偏差为-2.4,即-4.1%)具有良好的一致性。然而,对于二氧化碳(VCO(2),mL.min(-1))的测量值,一致性较差[-280 至 1394](平均偏差为 671,即 20.3%)。在最大运动强度下,Cortex 系统测量的 VCO(2)也显著高于 Cosmed 系统(p<0.01)。由于较高的测量 CO(2)浓度,使用 Cortex 设备计算得出的呼吸交换率(RER)值显著较高(p<0.01)。总之,Cosmed K4b(2)和 Cortex Metamax II 系统之间在大多数呼吸测量方面具有令人满意的一致性;然而,在这两个系统之间,VCO(2)和计算得出的 RER 测量值之间的一致性较差。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验