Microwave ablation with loop antenna: in vivo porcine liver model.

作者信息

Shock Sarah A, Meredith Kenneth, Warner Thomas F, Sampson Lisa A, Wright Andrew S, Winter Thomas C, Mahvi David M, Fine Jason P, Lee Fred T

机构信息

Department of Radiology, University of Wisconsin Hospitals and Clinics, E3/311 Clinical Science Center, 600 Highland Ave, Madison, WI 53792, USA.

出版信息

Radiology. 2004 Apr;231(1):143-9. doi: 10.1148/radiol.2311021342.

Abstract

PURPOSE

To determine the effectiveness of tissue ablation with a loop microwave antenna in various configurations in porcine liver tissue.

MATERIALS AND METHODS

Microwave energy was applied for 7 minutes at 60 W in six porcine livers (mean weight, 68.2 kg) by using single (n = 7) or dual 2.7-cm loop microwave probes in parallel (n = 9) or orthogonal (n = 9) configurations. Volume, diameter, shape, and temperature of the zone of necrosis and the presence of viable tissue inside the loop were determined and compared by means of factorial analysis of variance.

RESULTS

Mean lesion volume and maximum diameter, respectively, were 32.2 cm(3) +/- 14.4 (SD) and 4.6 cm +/- 1.4 for lesions ablated with parallel probes (parallel lesions), 29.5 cm(3) +/- 8.1 and 4.3 cm +/- 0.6 for lesions ablated with orthogonal probes (orthogonal lesions), and 6.4 cm(3) +/- 1.9 and 3.4 cm +/- 0.62 for lesions ablated with single probes (single lesions) (P <.05, single vs parallel and orthogonal lesions). Mean minimum diameter was greatest for orthogonal lesions (3.5 cm +/- 0.53; P =.017, parallel vs orthogonal lesions). Orthogonal lesions had the highest mean internal temperature (97.2 degrees C) versus parallel (91.9 degrees C) and single (60.0 degrees C) lesions. All orthogonal lesions heated to 60 degrees C in comparison to eight of nine parallel and four of seven single lesions. The mean time to reach 60 degrees C was shortest for orthogonal lesions (93.3 seconds) versus parallel (123.8 seconds) and single (263.0 seconds) lesions. Orthogonal lesions were the most spherical. Viable tissue was present in the center of five of seven single, six of nine parallel, and zero of nine orthogonal lesions.

CONCLUSION

Loop microwave antennas allow precise control and effective ablation of targeted tissue, particularly in the orthogonal configuration.