Abstract

PURPOSE

To evaluate the 3- to 4-year effects of the Implantable Contact Lens (ICL) on the corneal endothelium.

SETTING

Multicenter study.

METHODS

Noncontact specular microscopy was performed as a subgroup study in a Phase III U.S. Food and Drug Administration clinical trial. Endothelial cell images were collected in the central region of the cornea before surgery and 3, 12, 24, and 36 months after surgery, with a few at 48 months. The images were recorded and analyzed later by a central reading center. The cell density, coefficient of variation, and percentage of hexagonal cells were determined.

RESULTS

The cumulative endothelial cell loss was between 8.4% and 8.9% over the first 3 years and between 8.4% and 9.5% over the first 4 years depending on the method of calculation. The cell loss between baseline and 3 months was 2.1%; 3 months and 1 year, 0.9%; 1 year and 2 years, 2.3%; 2 years and 3 years, 3.2%; and 3 years and 4 years, -0.1%. The coefficient of variation decreased over the course of the study, and the proportion of cases with hexagonal cells increased slightly.

CONCLUSIONS

The cell loss between 1 year and 3 years in the absence of an increase in the coefficient of variation and/or a decrease in the percentage of hexagonal cells is most readily explained by prolonged corneal remodeling following the surgical procedure rather than ongoing cell loss. The cell loss observed between 3 years and 4 years (0.1% gain) was negligible. Regardless of the cause of the change in endothelial cell density over the first 3 years, the available 4-year data suggest there was no ongoing chronic loss.

摘要

目的

评估可植入式隐形眼镜(ICL)对角膜内皮细胞的3至4年影响。

设置

多中心研究。

方法

在美国食品药品监督管理局的一项III期临床试验中,作为亚组研究进行非接触式角膜内皮显微镜检查。在手术前以及手术后3、12、24和36个月(少数为48个月)收集角膜中央区域的内皮细胞图像。图像随后由中央阅片中心进行记录和分析。测定细胞密度、变异系数和六角形细胞百分比。

结果

根据计算方法,前3年累积内皮细胞损失在8.4%至8.9%之间,前4年在8.4%至9.5%之间。基线至3个月时细胞损失为2.1%;3个月至1年为0.9%;1年至2年为2.3%;2年至3年为3.2%;3年至4年为-0.1%。在研究过程中变异系数降低,六角形细胞比例略有增加。

结论

在变异系数未增加和/或六角形细胞百分比未降低的情况下,1至3年的细胞损失最容易用手术后角膜长期重塑来解释而非持续的细胞损失。3至4年观察到的细胞损失(增加0.1%)可忽略不计。无论前3年内皮细胞密度变化的原因如何,现有的4年数据表明不存在持续的慢性损失。