Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose.

作者信息

Sigal-Cinqualbre Anne B, Hennequin Remi, Abada Hicham T, Chen Xiaoyan, Paul Jean-François

机构信息

Department of Radiology, CT Unit, Hôpital Marie-Lannelongue, 133 Avenue de la Résistance, 92350 Le Plessis-Robinson, France.

出版信息

Radiology. 2004 Apr;231(1):169-74. doi: 10.1148/radiol.2311030191. Epub 2004 Feb 27.

Abstract

PURPOSE

To assess the feasibility of low-kilovoltage (ie, 80-kV) chest computed tomography (CT) protocols for adults and the effect of such protocols on image quality and iodine dose.

MATERIALS AND METHODS

Preliminarily, 90 patients (30 women, 60 men; mean age, 59 years) requiring contrast material-enhanced chest CT were randomly assigned to one of three protocol groups: protocol A, with use of 80 kV and 135 mAs; protocol B, with use of 80 kV and 180 mAs; or the standard protocol, with use of 120 kV and 90 mAs. Contrast material injection protocols were standardized in all groups. Image noise was calculated and plotted against patient weight. Subsequently, another 52 consecutive patients (11 women, 41 men; mean age, 57 years) were assigned to one of the protocols according to their weight: Patients weighing less than 60 kg were assigned to protocol A; patients weighing 60-75 kg, to protocol B; and patients weighing more than 75 kg, to the standard protocol. Two readers evaluated the CT images qualitatively by using a five-point scale. Statistical analyses were performed by using analysis of variance, kappa, and Fisher exact tests.

RESULTS

In the preliminary study, the mean image noise values with protocols A (24 HU) and B (20 HU) were significantly higher (P <.001) than that with the standard protocol (12 HU). With protocols A and B, in the patients weighing more than 60 kg and more than 75 kg, respectively, the noise increased exponentially with patient weight. In the subsequent study, qualitative analysis revealed no significant difference between the low-kilovoltage examinations and the standard examination. Compared with use of the standard protocol, use of protocols A and B resulted in the iodine-based contrast material dose being reduced by 54% and 39%, respectively.

CONCLUSION

Weight-adapted low-kilovoltage contrast-enhanced chest CT examinations can be routinely performed with 80 kV. Use of these protocols results in good diagnostic image quality and makes it possible to reduce contrast material use by more than 50%.

摘要

目的

评估低千伏(即80 kV)胸部计算机断层扫描(CT)方案用于成人的可行性,以及此类方案对图像质量和碘剂剂量的影响。

材料与方法

初步研究中,90例需要进行对比剂增强胸部CT的患者(30名女性,60名男性;平均年龄59岁)被随机分配至三个方案组之一:方案A,采用80 kV和135 mAs;方案B,采用80 kV和180 mAs;或标准方案,采用120 kV和90 mAs。所有组的对比剂注射方案均标准化。计算图像噪声并针对患者体重进行绘制。随后,另外52例连续患者(11名女性,41名男性;平均年龄57岁)根据其体重被分配至其中一个方案:体重小于60 kg的患者被分配至方案A;体重60 - 75 kg的患者被分配至方案B;体重超过75 kg的患者被分配至标准方案。两名阅片者使用五点量表对CT图像进行定性评估。采用方差分析、kappa检验和Fisher精确检验进行统计分析。

结果

在初步研究中,方案A(24 HU)和方案B(20 HU)的平均图像噪声值显著高于标准方案(12 HU)(P <.001)。采用方案A和方案B时,体重超过60 kg和超过75 kg的患者中,噪声随患者体重呈指数增加。在随后的研究中,定性分析显示低千伏检查与标准检查之间无显著差异。与使用标准方案相比,使用方案A和方案B分别使基于碘的对比剂剂量降低了54%和39%。

结论

采用80 kV可常规进行根据体重调整的低千伏对比增强胸部CT检查。使用这些方案可获得良好的诊断图像质量,并有可能将对比剂用量减少50%以上。