Effect of a facemask leak on aerosol delivery from a pMDI-spacer system.

作者信息

Esposito-Festen J E, Ates B, van Vliet F J M, Verbraak A F M, de Jongste J C, Tiddens H A W M

机构信息

Department of Pediatrics, Division of Respiratory Medicine, Erasmus MC-Sophia Children's Hospital, Rotterdam, The Netherlands.

出版信息

J Aerosol Med. 2004 Spring;17(1):1-6. doi: 10.1089/089426804322994406.

Abstract

The objective of this investigation was to study the relation between size and position of a mask leak on spacer output and lung dose. An upper-airway model (SAINT model, Erasmus MC) was connected to a breathing simulator. Facemasks with leaks ranging between 0 and 1.5 cm(2) were examined. Leaks were located close to the nose or close to the chin. During simulated breathing, 200 microg budesonide (Pulmicort, AstraZeneca) was delivered to the model via NebuChamber (AstraZeneca) with facemask. Spacer output and lung dose were measured by placing a filter between spacer and facemask or between model and breathing simulator, respectively. Budesonide trapped on the filter was quantified by means of HPLC, and expressed as percentage of the nominal dose. Mean spacer output doses for the nose position were 50, 38, 28, 12, 10, 6, and 0%, and for the chin position were 50, 40, 31, 11, 9, 4, and 0% for leaks of 0, 0.05, 0.1, 0.16, 0.2, 0.3, and larger than 0.4 cm(2), respectively. Mean lung doses for the nose position were 10, 8, 6, 3, 3, 1, 0, 0, 0, and 0%, and for the chin position were 10, 9, 8, 6, 6, 5, 1, 1, 0, and 0% for leaks of 0, 0.05, 0.1, 0.16, 0.2, 0.3, 0.4, 0.5, 1, and 1.5 cm(2). Efficiency of a pMDI-spacer facemask strongly depends on the size of a facemask leak. Spacer output did not depend on the position of the leak. Lung dose was higher for leaks near the chin than for leaks near the nose.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验