Wetzel Stephanie J, Guttman Charles M, Flynn Kathleen M
Polymers Division, National Institute of Standards and Technology, 100 Bureau Dr. Stop 8541, Gaithersburg, MD 20899, USA.
Rapid Commun Mass Spectrom. 2004;18(10):1139-46. doi: 10.1002/rcm.1462.
Although electrospray sample deposition in matrix-assisted laser desorption/ionization (MALDI) mass spectrometry (MS) sample preparation increases the repeatability of both the MALDI signal intensity and the measured molecular mass distribution (MMD), the electrospray sample deposition method may influence the apparent MMD of a synthetic polymer. The MMDs of three polymers of differing thermal stability, polystyrene (PS), poly(ethylene glycol) (PEG), and poly(propylene glycol) (PPG), were studied by MALDI time-of-flight (TOF) MS as the electrospray deposition voltage was varied. The MMDs obtained using the electrospray deposition method were compared with those obtained for hand-spotted samples. No change was observed in the measured polymer MMD when the electrospray deposition voltage was varied in the analysis of PS, but those of PEG and PPG changed at higher electrospray voltages due to increased ion fragmentation. It was also shown that the fragmentation in the hand-spotted samples is dependent on the matrix used in sample preparation.
尽管在基质辅助激光解吸/电离(MALDI)质谱(MS)样品制备中采用电喷雾样品沉积可提高MALDI信号强度和所测分子量分布(MMD)的重复性,但电喷雾样品沉积方法可能会影响合成聚合物的表观MMD。随着电喷雾沉积电压的变化,通过MALDI飞行时间(TOF)MS研究了三种热稳定性不同的聚合物,即聚苯乙烯(PS)、聚乙二醇(PEG)和聚丙二醇(PPG)的MMD。将采用电喷雾沉积方法获得的MMD与手动点样样品的MMD进行了比较。在PS分析中改变电喷雾沉积电压时,所测聚合物MMD未观察到变化,但由于离子碎片化增加,在较高电喷雾电压下PEG和PPG的MMD发生了变化。还表明,手动点样样品中的碎片化取决于样品制备中使用的基质。